121
BM5Vet Pro
User’s Manual
③
Temperature difference value display
④
Temperature unit display
Temperature Settings
A.
Temp menu
MENU
Description
Available Settings
A-1.
Alarm
Temp Alarm Settings menu
A-1-1.
PARAMETER ALARM
LIMIT
TEMP1, TEMP2, DELTA TEMP Parameter
Alarm level, Action setup menu
Settings range from 0
℃
to 50.0
℃
/
32°F to 122°F.
A-2.
DISPLAY OPTION
Temp waveform display setting
A-2-1.
DELTA DISPLAY
It can set whether to display two
temperature differences as a value.
ON/ OFF
A-2-2.
TEMP ORDER
It can set the order of the temperature
differences.
TEMP1
–
TEMP2
TEMP2
–
TEMP1
A-2-3.
2 CHANNEL VIEW
Whether to use 2 channels.
If only 1 channel is used, set it to OFF.
When using 1 channel, which channel
to use is determined by TEMP ORDER
ON/ OFF
Содержание BM5VET
Страница 22: ...22 BM5Vet Pro User s Manual Left Side 1 Printer...
Страница 30: ...30 BM5Vet Pro User s Manual How To Replace The Battery Please assemble and replace as shown below...
Страница 68: ...68 BM5Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Страница 155: ...155 BM5Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Страница 195: ...195 BM5Vet Pro User s Manual ACCVENT COUPLET IRREGULAR PAUSE RONT VBRADY SHORTRUN PVC...
Страница 219: ...219 BM5Vet Pro User s Manual SYSTEM LOWBATTERY...
Страница 225: ...225 BM5Vet Pro User s Manual less than minus number percent plus or minus...