33
Symptom
Possible cause
Solution
Droplets left inside the tip
Unsuitable tip
Use original tips
Leakage or pipetted
Non uniform wetting
Attach new tip
volume too small
of the plastic
Tip incorrectly attached
Attach firmly
Unsuitable tip
Use original tips
Foreign particles
Clean the tip cone,
between tip and cone
attach new tip
Instrument
Clean and grease
contaminated
the piston and the
tip cone
Insufficient amount of
Grease accordingly
grease on a piston and
o-ring
Pipettor out of
Instrument damaged
Return to your
established specifications
Biohit service
representative
for servicing
Pipettor blocked, aspirated
Liquid has penetrated
Clean and grease
volume too small
tip cone and dried
the piston and
the tip cone
Tip ejector jammed or
Tip ejector collar has
Remove and clean
moves erratically
been contaminated
ejector collar and
tip cone
Continuous error
Instrument damaged
Return to your
messages
Biohit service
representative
for servicing