32
12. Position the tip to aspirate and press the START button
13. Position the tip to dispense and press the START button
Note: After recalibration the display will show the operating mode symbol
followed with letters CAL to indicate that the mode and the volume in question
is recalibrated to give out the new volume.
Note: Changing the mode will reset the recalibration. However, it is possible to
store the recalibrated mode to the storage location in PROG mode as
described in paragraph 8.
12. Trouble shooting
The Biohit
e
LINE pipettors have a onboard monitoring program to control the
performance of each pipetting action. If any error message appears to the
display it means that the pipettor has been unable to perform the attempted
action properly. In case of an error the display will show Error and PRESS Ent.
message. To clear the message please follow these steps:
1.
As this procedure will reset the pipettor and the tip ejector please first
remove the tip by hand.
2.
Recharge the pipettor for 15 minutes.
3.
Clear the error message by pressing . The display will show RESET
and PRESS TIP.
3.
Reset the pipettor by pressing either of the tip ejector buttons .
Note: Repeated occurrence of error messages indicates an internal error
causing failure of the
e
LINE to complete the execution of the pipetting. The
e
LINE therefore needs to be returned to the nearest Biohit Service Centre or
your local Biohit service representative for repair.