5000: Unscrew the green tip cone (4) anti-clockwise and remove it.
Release the black tip ejector assembly from the tip cone by turning it anti-
clockwise and then pull it out. Remove the tip cone filter if fitted and
continue from Step 4.
3.
Unscrew the tip cone holder (3) anti-clockwise and remove the tip cone
holder (3), tip cone (4) and spring (5). Remove the tip cone filter if fitted.
4.
Unscrew the exposed piston (6) anti-clockwise and remove it.
Cleaning:
To clean the exposed parts use Proline
®
Biocontrol, ethanol (70%), isopropanol
(60%) or mild detergent and soft lint-free cloth. The interior of the tip ejector
collar and tip cone can be cleaned with a cotton swab. Rinse with distilled
water if needed. Let the parts dry. Grease the piston thinly with the grease
provided.
Decontaminate:
For complete decontamination place the
tip ejector collar (2), tip cone holder (3),
tip cone (4), spring (5) and piston (6) into
a beaker containing Biohit Proline
®
Biocontrol and leave for at least 30
minutes (Fig. 9), rinse the parts with
distilled water.
Note: The corresponding parts in the
e
5000 model are the green tip cone (4),
the black tip ejector assembly and the
piston (6).
Let the parts dry. Grease the piston thinly
with the grease provided.
Fig. 9.
Autoclaving:
The tip ejector collar (2), tip cone holder (3), tip cone (4), spring (5) and piston
(6) of the
e
LINE pipettor can be steam autoclaved (121
O
C, 1 bar overpressure
for 20 minutes), see Fig. 8. These parts can be autoclaved as one unit or
separately as individual parts. It is also possible to clean the parts and grease
the piston prior to autoclaving.
Note: The autoclavable parts in the
e
5000 model are the green tip cone (4), the
black tip ejector assembly and the piston (6).
Note: Avoid excess grease. Use only the grease provided with the pipettor.
Note: Before reassembling check that no lint or particles are on the surface of
the piston.