6.2. Reverse pipetting (
rP
)
A selected volume plus and excess is aspirated into the tip. After dispensing
the selected volume the excess volume remains in the tip and is discarded.
1.
Press to display PIPET
2.
Scroll with until the symbol rP is displayed.
3.
Confirm by pressing
4.
Select the desired pipetting volume by using
Note: When holding down the volume display starts to scroll faster.
5.
Press to confirm selection and to display the speed in
or
press and continue from Step 10.
6.
Select the speed in with (1= Min and 9 = Max)
7.
Press or to confirm selection and to display the speed out
8.
Select the speed out with (1= Min and 9 = Max)
9.
Press or to confirm selections
10. Position the tip to aspirate and press the START button
11. Position the tip to dispense. Press the START button
12. Position the tip to discard excess liquid and press the
START button twice
Step 10
Step 11 Step 12
Note: It is also possible to continue to aspirate and dispense the same volume
without the empty function. To continue, keep the START button depressed
during dispensing and within one second the direction arrow will change.
Continuing to hold the START button down, place the tip into the liquid again
and then release the START button to aspirate the next sample.
14