Reassembling:
1 .
Screw the piston (6) clockwise into position.
2.
e
10,
e
120,
e
300,
e
1000: Place the spring (5) around the piston.
e
5000: Attach the green tip cone by screwing it clockwise. Place the tip
ejector assembly to the tip cone and attach by turning it clockwise.
Replace the tip cone filter with the tool provided and continue from Step
5.
3.
Push the tip cone (4) into position and attach it by screwing the tip cone
holder (3) back. Replace the tip cone filter with the tool provided.
4.
Attach the tip ejector collar (2) by screwing it clockwise.
5.
Replace the battery pack (1). Reset the pipettor and press the start button
several times to test the movement. Test the tip ejection operation.
Note: When reassembling the pipettor do not over tighten the parts.
Note: It is always necessary to check the performance of the pipettor after in-
house service or maintenance (see paragraph 11).
10.3. Battery replacement
The design of the Biohit
e
LINE pipettors enables fast and easy battery
replacement. The battery replacement pack includes 4 NiMH batteries fixed to
the battery cover. If the batteries do not hold a sufficient charge for proper
operation follow these steps for replacement of the battery pack.
Note: The battery pack fits in to the compartment one way only. If the pack is
inserted incorrectly the battery cover will not close properly.
Note: Only use battery packs supplied by the manufacturer. The use of any
other type of battery will immediately invalidate the warranty!
Note: Dispose the used battery pack in accordance of your local regulations.
28
1.
Hold the pipettor in a horizontal position
with the display facing up.
2.
Press the two charging contacts
simultaneously to release the battery
pack. (Fig. 10.)
3.
Place the new battery pack into the
battery compartment and close the
battery cover by pressing it lightly.
Fig. 10.