18
5.
Press or to confirm the selection and to display the aliquots
6.
Select the desired number of aliquots with
7.
Press to confirm selection and to display the speed in
or
press and continue from Step 12.
8.
Select the speed in with (1= Min and 9 = Max)
9.
Press or to confirm selection and to display the speed out
10. Select the speed out with (1= Min and 9 = Max)
11. Press or to confirm selections
12. Position the tip to aspirate and press the START button . The display
shows PRE OUT and arrow down to indicate the reset function.
13. Position the tip to discard priming excess and press the start button .
14. To dispense, position the tip, press START button and repeat until the
cycle is complete.
15. Finally position the tip to discard any remaining excess and press the
START button twice .
Step 12 Step 13 Step 14 Step 15
Note: It is also possible to continue to aspirate and dispense the same volume
without the empty function. To continue, keep the START button pressed during
the last dispensing and within one second the direction arrow will change.
Keeping the START button depressed, place the tip into the liquid again and
then release the START button to aspirate the next sample.
▲
▼
▼
▼