9
1.
Check that the battery protector is removed and the battery cover is
properly closed.
2.
Place the pipettor into the Charging unit. Make sure that the metal
contacts at the rear of the pipettor are properly fitted to the tracks of the
metal contacts in the top of the Charging unit.
3.
When charging, the battery symbol will be displayed with the + and –
symbols flashing. The charging time for completely discharged batteries
is 4 hours.
4.
When the battery is full the battery symbol disappears from the display. If
display shows RESET and PRESS TIP, press the tip ejection button to
reset the pipettor, the pipettor will then be ready for use.
Note: Never reset the pipettor with the tip attached as the attached tip can
induce an error in the home position determination of the piston and tip ejector
collar!
Note: When the pipettor is not in use it is recommended to keep the pipettor in
the Charging unit. This ensures that the battery is kept charged and the pipettor
is ready for use.
4.2. Charging through direct charging system
The AC-adaptor supplied with the Biohit
e
LINE pipettor
or Charging Stand can be used for direct charging of the
Biohit
e
LINE electronic pipettor. The socket for the AC-
adaptor is placed at the back of the pipettor (Fig. 6).
1.
Before connecting the AC-adaptor to the electrical
outlet, check that its voltage settings corresponds
to the voltage settings of your electrical outlet.
2.
Check that the battery protector is removed and
the battery cover is properly closed.
3.
Connect the AC-adaptor to the electrical outlet as
well as to the socket at the back of the pipettor.
4.
The charging time is 4 hours for a completely
discharged battery.
Note: It is possible to continue pipetting while the
e
LINE
pipettor is charged through the AC-adaptor (Fig. 7). With
fully discharged battery a 5 minute initial charging is
required before continuing the pipetting.
Fig. 7.
Fig. 6.
Содержание eLINE 73002X