17
11. Press the to change the aspiraton/dispensing direction.
12. Position the tip to dispense. Press the START button
13. When the tip is empty (volume reaches 0
µ
l), within one second the display
will show E. Position the tip to empty the tip with blow-out and press the
START button twice.
Step 12 Step 13
7. MULTI
modes
The mode range
MULTI
includes 6 different operating modes.
7.1. Multi-dispensing (
d
)
The pipettor performs repetitive dispensing of a selected volume. The sum of
the dispensing aliquots and an automatically selected excess volume is
aspirated into the tip. The excess volume is needed to ensure equal operating
conditions for each dispensing step.
1.
Press to display
MULTI
2.
Scroll with until the symbol d is displayed.
3.
Confirm by pressing
4.
Select the desired dispensing volume by using
Note: When holding down the the volume display starts to scroll faster.
Symbol
Description
d
Multi-Dispensing
dd
Diluting
dd +
MIXING
Diluting with mixing function
Sd
Sequenced dispensing
SA
Multi-aspirating
Ad
Automatic multi-dispensing with selected interval
▼
▼