6.3. Pipetting with cycle counter (
P
COUNT
)
The
e
LINE performs pipetting with blow-out and the cycle counter function. The
cycle counter counts up to 99 pipettings and then returns back to 0. It is
possible to choose the cycle number to begin with. It is also possible to change
the cycle number or reset the counter during the pipetting by using the arrow
keys.
1.
Press to display PIPET
2.
Scroll with until the symbol P and
COUNT
is displayed.
3.
Confirm by pressing
4.
Select the desired pipetting volume by using the
Note: When holding down the volume display starts to scroll faster.
5.
Press or to confirm selection.
6.
Cycle counter will show 1. Use to select the sample number you
want to begin with (0-99).
7.
Press to confirm the selection and to display the speed in
or
press and continue from Step 12.
8.
Select the speed in with (1= Min and 9 = Max)
9.
Press or to confirm selection and to display the speed out
10. Select the speed out with (1= Min and 9 = Max)
11. Press or to confirm selections
12. Position the tip to aspirate and press the START button
13. Position the tip to dispense and press the START button . The tip is
emptied with blow-out and the piston returns automatically back to the
home position. The pipettor is ready for next pipetting.
Step 12 Step 13
Note: To change the cycle number or to reset the counter during the pipetting,
press the to display the desired cycle number and confirm with .
15
Содержание eLINE 73002X
Страница 36: ...36 Notes...