12
5.2. Speed adjustment
There are 9 speeds available for both dispensing and aspiration. The speed is
indicated in the bottom left hand corner of the display. The default speed is 5
for all operating modes.
Speed can be adjusted during the programming by completing the setting
changes with the select key. In every operating mode the speed adjustment is
the last programming step. After adjusting the “speed out” the programming is
completed with enter or select key and the pipettor is then ready for operation.
Note: The P
MANUAL
mode includes 5 different dispensing and aspirating
speeds.
6. PIPET
modes
The mode range PIPET includes 5 different operating modes.
6.1. Pipetting mode (
P
) and pipetting with mixing (
P+
MIXING
)
In pipetting mode (P) the
e
LINE performs pipetting with blow-out. In pipetting
with mixing mode (P
+MIXING
) pipetting is followed by a mixing function.
1.
Press to display PIPET
2.
Scroll with until the symbol P or P+
MIXING
is displayed.
3.
Confirm by pressing
4.
Select the desired pipetting volume by using the
Note: When holding down the the volume display starts to scroll faster.
5.
Press to confirm selection and to display the speed in
or press and continue from Step 10.
Symbol Description
P
Pipetting with automatic blow-out
rP
Reverse pipetting without blow-out
P +
MIXING
Pipetting with blow-out followed with mixing function
P
COUNT
Pipetting with blow-out and displayed cycle counter
P
MANUAL
Manual pipetting with on-line volume display
Содержание eLINE 73002X