2.4.3
Signal interface
1.
Solder the litz wires to the connector as required for the desired
operating mode. The available operating modes depend on the
device variant; see chapter 3.
ATTENTION
Free, unassigned pins must not be wired.
NOTE
Connect system supply voltage ground to protective ground.
2.
Push the shield back and fix with a cable tie.
3.
Insert two threaded bolts (fig. 2-7) into the connector shell.
4.
Place the connector into the connector shell.
5.
Fasten the cable and the shield to the connector shell with screws,
providing for strain relief.
ATTENTION
Ensure good electrical contact between the shield and the con-
nector shell.
Connect the shield on both ends.
6.
Assemble the two parts of the connector shell with two screws.
7.
Fasten the connector to the front panel (item 23) with screws.
DANGER
All signal connections must be definitely isolated from mains. The
voltage towards ground must not exceed 60 VDC or 25 VAC. All
signal circuits are internally grounded via a 1 Mohm bleed resistor.
Fig. 2-7 Signal connector
assembly – device end
Installation
WDP3-01X Ident. no.: 00441109760
2-9