Specific functions of BML BM1/BM2 (continued)
Calibration errors
Fig. 8-9: Calibration error
The errors in Tab. 8-4 can occur during calibration:
Error description
Troubleshooting
Speed too low or no movement detected.
The speed does not agree with the set speed range.
►
Check the speed range and if necessary adjust it in the settings.
Speed too high.
Communication error. Check the connection
and restart the Calibration Wizard.
►
Check the connection on the sensor module and USB adapter and
restart the Calibration Wizard.
Insufficient working memory. Try closing some
programs and restarting the Calibration Wizard.
►
Close as many programs as possible and restart the Calibration
Wizard.
Unknown error. Restart the Calibration Wizard.
►
Check the connection on the sensor module and USB adapter and
restart the Calibration Wizard.
Tab. 8-4: Calibration error BML BM1