AquaTrans™ AT600

User’s manual

Содержание Panametrics AquaTrans AT600

Страница 1: ...panametrics com BH001C11 Rev F AquaTrans AT600 User s manual...

Страница 2: ......

Страница 3: ...more registered trademarks of Baker Hughes Company and its subsidiaries in one or more countries All third party product and company names are trademarks of their respective holders AquaTrans AT600 Pa...

Страница 4: ...ii no content intended for this page...

Страница 5: ...emote and on site support needs To complement our broad portfolio of industry leading solutions we offer several types of flexible and scalable support services including Training Product Repairs Serv...

Страница 6: ...pact health and the environment In order to avoid the dissemination of those substances in our environment and to diminish the pressure on the natural resources we encourage you to use the appropriate...

Страница 7: ...T transducers with an RG62 cable 17 Chapter 3 Wiring the AT600 electronics 3 1 Wiring diagram 19 3 2 Wiring the line power 21 3 3 Wiring the transducers 23 3 4 Wiring the system ground 23 3 5 Wiring t...

Страница 8: ...59 4 7 4 3 Setting the pipe lining 61 4 7 5 Programming the transducer when clamp on selected 62 4 7 5 1 Entering a standard transducer 62 4 7 5 2 Entering a special transducer 64 4 7 6 Programming t...

Страница 9: ...4 3 Display languages 144 A 4 4 Product models 144 Appendix B Specifications B 1 Service record 145 B 2 Data entry 145 B 3 Initial settings 146 B 4 Diagnostic parameters 147 Appendix C Specifications...

Страница 10: ...Contents viii AquaTrans AT600 User s manual no content intended for this page...

Страница 11: ...including water wastewater steel campus energy and others The AT600 utilizes a new electronics platform and industrial design to make it extremely simple to install and use in the field So easy to us...

Страница 12: ...rtain number of acoustic pulses and then as a receiver for an identical number of pulses see figure 2 and figure 3 below The time interval between transmission and reception of the ultrasonic signals...

Страница 13: ...ansducer system on page 8 WARNING The AT 600 flow transmitter can measure the flow rate of many fluids some of which are potentially hazardous The importance of proper safety practices cannot be overe...

Страница 14: ...able in a variety of configurations to meet your needs so the packing list will vary for each system As an example a typical packing list is 1 One AT600 electronics enclosure 2 Two clamp on fixtures 3...

Страница 15: ...aluminum NEMA type 4X IP67 enclosure suitable for indoor or outdoor use See figure 5 below for the mounting dimensions and weight of the AT600 electronics enclosure Figure 5 Mounting the AT600 electro...

Страница 16: ...00 mounting base Figure 6 The AT600 mounting base 1 6 Calculating the transducer spacing Before installing the clamp on fixture s and transducers you must program the AT600 to calculate the required t...

Страница 17: ...ing sources of turbulence in the fluid such as valves flanges expansion joints elbows swirl and cavitation Locate the transducers on a common axial plane along the pipe see figure 8 below The transduc...

Страница 18: ...your specific configuration IMPORTANT See Sensor setup on page 56 to calculate the required transducer spacing A two traverse installation is recommended for most applications IMPORTANT If there is a...

Страница 19: ...Transducer frequency MHz Number of traverses Transducer spacing mm Number of fixtures mm inches 50 to 100 2 to 4 2 4 32 to 250 1 100 to 150 4 to 6 2 2 32 to 250 1 50 to 150 2 to 6 2 1 0 to 250 2 100...

Страница 20: ...Chapter 1 General installation instructions 10 AquaTrans AT600 User s manual no content intended for this page...

Страница 21: ...tallation Proceed as follows 1 Install the AT600 clamp on fixture with transducers on the pipe using two mounting straps a Choose a location with enough straight pipe run refer to Figure 7 on page 7 b...

Страница 22: ...13 below Figure 13 Set transducer spacing and apply couplant 5 See figure 14 below for an example of a completed installation In this drawing LSL is the lower specified limit and USL is the upper spe...

Страница 23: ...nsducers or cable on the pipe between the downstream pair of straps Use the bar on the second fixture to connect the two fixtures Then move the straps onto the second fixture and tighten the screws IM...

Страница 24: ...sducer spacing 0 to 250 mm or 0 to 320 mm traverses 1 fixtures 2 For a one traverse installation with a calculated transducer spacing of 0 to 250 mm for a 2 MHz transducer or 0 to 320 mm for a 1 MHz o...

Страница 25: ...ugh one side of the rail on the first fixture and inserted through the side of the rail on the second fixture c Install the upstream transducer into the second fixture and reconnect the transducer cab...

Страница 26: ...f this clamp on fixture Tighten the strap screws and verify that the straps remain on the ends of the fixture 6 Set the spacing between the two transducers to the value calculated by the meter as foll...

Страница 27: ...the BNC connector on the C RS or C PT transducer to the SMA connector on the AT6 transducer cable a BNC to SMA adapter is needed See figure 22 below and install the cable adapter on the transducer en...

Страница 28: ...Chapter 2 Clamp on fixture and transducer installation 18 AquaTrans AT600 User s manual no content intended for this page...

Страница 29: ...ions must be selected when ordering the AT600 Figure 23 Wiring diagram Attention European Customers To meet CE Mark requirements all cables must be installed as described in Wiring cable specification...

Страница 30: ...red cable specifications IMPORTANT Any unused cable glands must be plugged with the cable gland inserts provided with meter Figure 24 Recommended cable gland usage GLAND HOLE 5 GLAND HOLE 3 GLAND HOLE...

Страница 31: ...ed to the internal ground connection 3 Connect the neutral AC lead or the negative DC lead to L2 N and the AC line lead or the positive DC lead to L1 as shown in Figure 23 on page 19 IMPORTANT Do not...

Страница 32: ...600 electronics 22 AquaTrans AT600 User s manual Figure 25 Typical XMT1000 Labels Note This label snapshot is only for illustration purpose The information on actual label may differ owing to specific...

Страница 33: ...own in Figure 23 on page 19 3 Secure the cable gland IMPORTANT Be sure to insert all cable connectors straight into the PCB receptacles to avoid damaging the connector and or the receptacle 3 4 Wiring...

Страница 34: ...internally by the meter Do not connect an external 24V power supply to this circuit Prior to use the analog output must be set up and calibrated see Inputs outputs on page 41 When in meter is in conf...

Страница 35: ...can start or stop the totalizer functionality by toggling the gate switch To wire the Gate input refer to Figure 23 on page 19 and figure 29 below and complete the following steps 1 Verify that the ma...

Страница 36: ...Chapter 3 Wiring the AT600 electronics 26 AquaTrans AT600 User s manual no content intended for this page...

Страница 37: ...s for programming the AT600 flow meter to place it into service Before the AT600 can begin taking measurements at least the following menus must be programmed User preferences on page 37 Inputs output...

Страница 38: ...rs to exit from a specific option without saving unconfirmed data and enable users to highlight a specific window in the display option or to scroll through a list of options They are also used to cha...

Страница 39: ...ypad operation on page 28 and the following two LEDs Green The green LED is a system health indicator and it is on when the meter is operating without error Red The red LED is a system status indicato...

Страница 40: ...s in the displayed value From the display screen press either the or keys until the value is highlighted After the value is highlighted press to open the display decimal option Use the and keys to scr...

Страница 41: ...alizer screen proceed as follows From the display screen press either the or keys until the value is highlighted After the value is highlighted press to open the measurement type option The screen cha...

Страница 42: ...nd Sci Scientific Notation Press to select the value and then press again to confirm the selection or to cancel the selection From the display screen press either the or keys until the measurement typ...

Страница 43: ...hlighted Note If you selected time the available units are seconds minutes hours and days 4 3 4 Starting or stopping the totalizer measurement To start or stop totalizer measurements After the unit is...

Страница 44: ...alizer To reset the totalizer proceed as follows 4 4 Entering the main menu Figure 33 The main menu map From the display screen press either the or keys until the reset icon a partial circle with an a...

Страница 45: ...tial screen use the arrow keys to highlight the lock symbol and press The following screen opens Use the or keys to highlight display format and press The following screen opens Use the and arrow keys...

Страница 46: ...t a valid password The next section describes the access levels required to edit various parameters To view all parameters without editing select program review Figure 34 Program and program review me...

Страница 47: ...er preferences 4 5 1 Setting Note You can only change the meter tag and label data using Panametrics Vitality software To enter the programming menu on the display menu use the arrow keys to highlight...

Страница 48: ...check or change velocity flow units use the or arrow key to select units setting and press The following screen opens Under units setting menu use the or arrow key to select the unit that need to be...

Страница 49: ...ty and press The following screen opens Use the or arrow keys to highlight the desired density type and press Press again and a screen similar to the following opens Use the or key to select a specifi...

Страница 50: ...quired After all digits are correct press to save Then press x to return to the previous screen The default password is 1111 To turn the backlight OFF or ON use the or arrow key to select display and...

Страница 51: ...re correct press to save Then press x three times to return to the user preference screen To access the analog output menu on the initial screen highlight the lock symbol and press The following scree...

Страница 52: ...red output with the or arrow keys and press to enter the configuration menu Select measurement and press The following screen opens In the measurement menu select the type of analog output to be used...

Страница 53: ...mA level or percentage of scale to test the output linearity Select the desired option and press A screen similar to the following opens Read the analog output with a digital multimeter and enter that...

Страница 54: ...l screen highlight the lock symbol and press The following screen opens Select program and press The following screen opens Use the or key to select a specific digit then use the or key to change that...

Страница 55: ...d press to open the following screen In the measurement menu select the type of analog output to be used and press to return to the previous screen Use the or arrow keys to select pulse value and pres...

Страница 56: ...h pulse is PV VR 0 23 4 6 50 ms As the pulse time PT is defined as half the pulse duration then the correct PT 25 ms Program the AT600 frequency output with PV 0 23 liters pulse and PT 25 ms If your f...

Страница 57: ...ct test pulse and press The following screen opens Press to open a screen similar to the following Press again and the screen returns to the previous display Press x to return to the digital output me...

Страница 58: ...he previous screen The base value is the measurement that corresponds to a 0 Hz pulse The full value is the measurement value that corresponds to a full frequency pulse The full frequency is the maxim...

Страница 59: ...old hold the last good value Low display 0 Hz High display the full frequency Other If selected a screen similar to the following opens Enter the frequency value you want to display during an error si...

Страница 60: ...ed If desired you can repeat this procedure with several different frequencies After testing is complete press x to return to the digital output menu The alarm can be either an open circuit normally c...

Страница 61: ...status and press The screen returns to the previous display Use the or arrow keys to select alarm type and press A screen similar to the following opens You can choose from three alarm types Low the a...

Страница 62: ...hat digit as required After all digits are correct press to save Then press x to return to the previous screen To test the alarms output select test alarms and press The following screen opens Select...

Страница 63: ...4 6 4 1 Modbus Selecting the baud rate Selecting the Modbus address To set up Modbus select digital comm on the input output screen and press The following screen opens Press again and a screen simila...

Страница 64: ...hat digit as required After all digits are correct press to save Then press x to return to the previous screen To set the bits parity select bits parity and press A screen similar to the following ope...

Страница 65: ...screen and press The following screen opens Press again and a screen similar to the following opens To set the HART address select address and press A screen similar to the following opens Press agai...

Страница 66: ...om the measurement display highlight the lock symbol and press The following screen opens Select program and press The following screen opens Use the or key to select a specific digit then use the or...

Страница 67: ...cutoff and press The following screen opens Press again and a screen similar to the following opens Use the or key to select a specific digit then use the or key to change that digit as required Afte...

Страница 68: ...hanges the pipe ID Select wetted if need and press to return to the previous screen Select pipe and press The following screen opens In the sensor setup menu select pipe OD pipe ID or wall thickness a...

Страница 69: ...te Table 3 below lists the available pre programmed pipe materials Select pipe and press The following screen opens In the sensor setup menu select pipe OD pipe ID or wall thickness and press A screen...

Страница 70: ...ASS Flint glass CROWN GLASS Crown glass NYLON PLSTC Nylon plastic POLYE PLSTC Polyethylene POLYP PLSTC Polypropylene PVC PLSTC Polyvinyl chloride ACRYL PLSTC Acrylic plastic Select the appropriate pip...

Страница 71: ...similar to the following opens To set the lining thickness select lining thickness and press A screen similar to the following opens Press again and a screen similar to the following opens Use the or...

Страница 72: ...ing material Tar epoxy Tar epoxy Pyrex glass Pyrex glass Asbestos cement Asbestos cement Mortar Mortar Rubber Rubber Teflon Teflon PFTE Press again and a screen similar to the following opens Use the...

Страница 73: ...quired After all digits are correct press to save Then press x to return to the previous screen Transducer number Transducer type 10 CPT 0 5CPT 0 5 11 CPT 2 0 12 CPT 0 5 MT C PB 05 M 13 CPT 1 0 MT C P...

Страница 74: ...2 0 HT 38 CPB 1 0 39 CPB 1 0 MT 301 C RL 0 5 302 C RL 1 304 C RL 0 5 305 C RL 1 307 C RL 0 5 308 C RL 1 310 C RV 0 5 311 C RV 1 313 C RW 0 5 314 C RW 1 401 C RS 0 51 402 C RS 11 403 C RS 2 407 UTXDR 2...

Страница 75: ...revious screen Select wedge angle and press The following screen opens Press again and a screen similar to the following opens Use the or key to select a specific digit then use the or key to change t...

Страница 76: ...digits are correct press x to save Then press x to return to the previous screen Select time wedge and press The following screen opens Press again and a screen similar to the following opens Use the...

Страница 77: ...ng screen opens Select STD and press A screen similar to the following opens Press again and a screen similar to the following opens Use the or key to select a specific digit and change that digit as...

Страница 78: ...ducer and press The following screen opens Select SPEC and press A screen similar to the following opens In the special transducer menu select frequency and press The following screen opens Select the...

Страница 79: ...ain and a screen similar to the following opens Use the or key to select a specific digit then use the or key to change that digit as required After all digits are correct press to save Then press x t...

Страница 80: ...he previous screen Select fluid type and press The following screen opens If the fluid type is water select water and press to return to the previous screen If the fluid is not water select other and...

Страница 81: ...ecific digit then use the or key to change that digit as required After all digits are correct press to save Then press x to return to the previous screen Select fluid temperature in the sensor setup...

Страница 82: ...hanging the transducer spacing should only be completed according to instructions received from the factory Select path and press The following screen opens Press again Record this calculated transduc...

Страница 83: ...screen opens In the sensor setup menu select path length or axial length and press A screen similar to the following opens Press again Enter this path use in installing your transducers on the pipe T...

Страница 84: ...Chapter 4 Initial setup and programming 74 AquaTrans AT600 User s manual no content intended for this page...

Страница 85: ...ignal transit time NA NA Tdn Displays the downstream ultrasonic signal transit time NA NA DeltaT Displays the transit time difference between the upstream and downstream signals NA NA Up signal qualit...

Страница 86: ...locity exceeds the limits programmed in the program advanced error limits velocity low high menu Cause When the measured velocity exceeds the programmed velocity limit this error is displayed The erro...

Страница 87: ...sical installation of the system meets the recommended specifications it is possible that the fluid itself may be preventing accurate flow rate measurements The fluid being measured must meet the foll...

Страница 88: ...the correct data will be included in the documentation For other flowcells measure the pipe wall thickness and diameter with the same accuracy desired in the flow rate readings Also check the pipe fo...

Страница 89: ...CHAR 32 128 296 User Product descriptor for HART RW CHAR 16 140 140 320 User Product electronic serial number RW CHAR 16 148 328 User Product fixture serial number RW CHAR 16 150 336 User Product tran...

Страница 90: ...tocol type RO INT32 900 900 2304 Viewer Display language RW INT32 90A 2314 Viewer Display variable 2 type RW INT32 90C 2316 Viewer Display totalizer 1 type RW INT32 90E 2318 Viewer Display totalizer 2...

Страница 91: ...96 User Digital out 1 pulse error handling RW INT32 D46 3398 User Digital out 1 pulse time RW INT32 D50 D50 3408 User Digital out 2 pulse measurement type RW INT32 D52 3410 User Digital out 2 pulse te...

Страница 92: ...er HART dynamic variable index_2 RW INT32 1144 4420 Viewer HART dynamic variable index_3 RW INT32 1146 4422 Viewer HART dynamic variable index_4 RW INT32 1300 1300 4864 RO HART configuration change co...

Страница 93: ...6 User MultiK velocity 1 RW IEEE 32 bit 2042 8258 User MultiK velocity 2 RW IEEE 32 bit 2044 8260 User MultiK velocity 3 RW IEEE 32 bit 2046 8262 User MultiK velocity 4 RW IEEE 32 bit 2048 8264 User M...

Страница 94: ...2 bit 2100 2100 8448 User Pipe material RW INT32 2102 8450 User Lining material RW INT32 2104 8452 User XDR type RW INT32 2106 8454 User XDR frequency RW INT32 2108 8456 User XDR wedge type RW INT32 2...

Страница 95: ...228A 8842 RO Up amp disc RO IEEE 32 bit 228C 8844 RO Dn amp disc RO IEEE 32 bit 228E 8846 RO SNR on UP channel RO IEEE 32 bit 2290 8846 RO SNR on DOWN channel RO IEEE 32 bit 2292 8850 RO Time in buffe...

Страница 96: ...ad unique identifier associated with tag If the specified tag matches that of the meter it responds with the command 0 response 12 Read message Reads the message contained within the meter 13 Read tag...

Страница 97: ...condary tertiary and quaternary variables 54 Read device variable information Get device variable information 59 Write number of response preambles Sets the number of asynchronous preamble bytes to be...

Страница 98: ...cific response codes for login with password Command 129 0x81 Logout and save This command will save any changes and logout from the flow meter Table 14 Request data bytes for logout and save Table 15...

Страница 99: ...87 Read current user access right This command will read the current user access right Table 20 Request data bytes for read current user access right Code Class Description 0 Success No command specif...

Страница 100: ...ata bytes for sends new password Table 24 Response data bytes for sends new password Table 25 Command specific response codes for sends new password Byte Format Description None Code Class Description...

Страница 101: ...nit 2 Actual volumetric unit 3 Standard volumetric unit 4 Mass unit 5 Totalizer unit 6 Density unit 7 Pipe dimension 8 Thermal 9 Acceleration Byte Format Description 0 Unsigned 8 Group index 1 Velocit...

Страница 102: ...ting Table 33 Response data bytes for read backlight setting Byte Format Description 0 Unsigned 8 Density type 1 Actual density 2 Reference density Byte Format Description 0 Unsigned 8 Density type 1...

Страница 103: ...No command specific errors 1 5 Undefined 6 Error Device specific command error 7 127 Undefined Byte Format Description 0 Unsigned 8 Group index 1 Velocity unit 2 Actual volumetric unit 3 Standard volu...

Страница 104: ...Undefined 5 Error Too few data bytes received 6 Error Device specific command error 7 Error In write protect mode 8 15 Undefined 16 Error Access restricted 17 127 Undefined Byte Format Description 0...

Страница 105: ...for read analog measurement range value 8 15 Undefined 16 Error Access restricted 17 127 Undefined Byte Format Description 0 Unsigned 8 Backlight control switch 0 off 1 on 1 4 Unsigned 32 Display back...

Страница 106: ...lue Table 49 Command specific response codes for read loop current error handling 1 4 Float Upper range value 5 8 Float Lower range value Code Class Description 0 Success No command specific errors 1...

Страница 107: ...or set loop current zero Byte Format Description 0 Unsigned 8 Fixed current level 0 Exit fixed loop current 1 fixed 4 mA 2 fixed 20mA 3 fixed percentage of scale Byte Format Description 0 Unsigned 8 F...

Страница 108: ...vel units of milliamperes Code Class Description 0 Success No command specific errors 1 2 Undefined 3 Error Passed parameter too large 4 Error Passed parameter too small 5 Error Too few data bytes rec...

Страница 109: ...ct loop current mode or value 10 15 Undefined 16 Error Access restricted 17 31 Undefined 32 Error Busy 33 127 Undefined Byte Format Description 0 3 Float Loop current percentage units of percent Byte...

Страница 110: ...lue 5 8 Float Lower range value Byte Format Description 0 Unsigned 8 Upper and lower range values unit code 1 4 Float Upper range value 5 8 Float Lower range value Code Class Description 0 Success No...

Страница 111: ...s to read the digital output configuration Table 68 Request data bytes Byte Format Description 0 Unsigned 8 Analog output error handling 0 Low 1 High 2 Hold 3 Other value 1 4 Float Error value unit is...

Страница 112: ...tal output type 0 Off 1 Pulse 2 Frequency 3 Alarm Code Class Description 0 Success No command specific errors 1 Undefined 2 Error Invalid selection 3 4 Undefined 5 Error Too few data bytes received 6...

Страница 113: ...selection 3 4 Undefined 5 Error Too few data bytes received 6 Error Device specific command error 7 127 Undefined Byte Format Description 0 Unsigned 8 Channel number 1 2 Byte Format Description 0 Unsi...

Страница 114: ...read alarm configuration Byte Format Description 0 Unsigned 8 Channel number 1 2 Byte Format Description 0 Unsigned 8 Channel number 1 Unsigned 8 Measurement type 2 Unsigned 8 Alarm value unit 3 6 Flo...

Страница 115: ...tal configuration Byte Format Description 0 Unsigned 8 Channel number 1 2 1 Unsigned 8 Digital output type 0 Off 1 Pulse 2 Frequency 3 Alarm Byte Format Description 0 Unsigned 8 Channel number 1 2 1 U...

Страница 116: ...total 2 Unsigned 8 Pulse value unit 3 6 Float Pulse value 7 10 Unsigned 32 Pulse time unit is ms 11 Unsigned 8 Pulse error handling 2 Hold good value 4 Stop Byte Format Description 0 Unsigned 8 Channe...

Страница 117: ...Unsigned 8 Frequency value unit 3 6 Float Frequency base value 7 10 Float Frequency full value 11 14 Unsigned 32 Full frequency unit is Hz 15 Unsigned 8 Frequency error handling 0 Low 1 High 2 Hold 3...

Страница 118: ...error 7 Error In write protect mode 8 15 Undefined 16 Error Access restricted 8 127 Undefined Byte Format Description 0 Unsigned 8 Channel number 1 2 1 Unsigned 8 Measurement type 2 Unsigned 8 Alarm v...

Страница 119: ...ommand specific errors 1 Undefined 2 Error Invalid selection 3 4 Undefined 5 Error Too few data bytes received 6 Error Device specific command error 7 Error In write protect mode 8 15 Undefined 16 Err...

Страница 120: ...e 98 Request data bytes for read pipe material 3 4 Undefined 5 Error Too few data bytes received 6 Error Device specific command error 7 Error In write protect mode 8 15 Undefined 16 Error Access rest...

Страница 121: ...0xC3 Read sensor meter setup This command is to read the sensor meter setup Table 104 Request data bytes for read sensor meter setup Byte Format Description 0 3 Unsigned 8 Pipe material 4 7 Float Pip...

Страница 122: ...es for read transducer information Byte Format Description 0 3 Float Zero cutoff Code Class Description 0 Success No command specific errors 1 5 Undefined 6 Error Device specific command error 7 127 U...

Страница 123: ...CPB 2 0 HT 38 CPB 1 0 39 CPB 1 0 MT 301 C RL 0 5 302 C RL 1 304 C RL 0 5 305 C RL 1 307 C RL 0 5 308 C RL 1 310 C RV 0 5 311 C RV 1 313 C RW 0 5 314 C RW 1 401 C RS 0 5M 402 C RS 1M 403 C RS 2M 407 U...

Страница 124: ...ad fluid information Table 113 Request data bytes for read fluid information Table 114 Response data bytes for read fluid information 6 Error Device specific command error 7 127 Undefined Byte Format...

Страница 125: ...fic errors 1 5 Undefined 6 Error Device specific command error 7 127 Undefined Byte Format Description 0 Unsigned 8 Pipe size unit 1 4 Float Pipe OD value 5 8 Float Pipe ID value 9 12 Float Pipe WT va...

Страница 126: ...pe lining attribute Byte Format Description 0 3 Unsigned 32 Pipe material 4 7 Float Pipe sound speed Byte Format Description 0 3 Unsigned 32 Pipe material 4 7 Float Pipe sound speed Code Class Descrip...

Страница 127: ...setup Code Class Description 0 Success No command specific errors 1 4 Undefined 5 Error Too few data bytes received 6 Error Device specific command error 7 Error In write protect mode 8 15 Undefined 1...

Страница 128: ...CPT 0 5 MT C PB 05 M 13 CPT 1 0 MT C PB 10 M 14 CPT 2 0 MT C PB 20 M 15 CPT 0 5 HT 16 CPT 1 0 HT 17 CPT 2 0 HT 18 CPS 0 5 19 CPSM 2 0 20 CTS 1 0 21 CTS 1 0 HT 22 CTS 2 0 23 C LP 40 HM 24 C LP 40 NM 25...

Страница 129: ...type 12 15 Unsigned 32 Transducer wedge angle 16 19 Unsigned 32 Transducer wedge SOS 20 23 Unsigned 32 Transducer Tw Byte Format Description 0 3 Unsigned 32 Transducer type 0 Other 4 7 Unsigned 32 Tr...

Страница 130: ...d 8 Transducer traverse 1 4 Unsigned 32 Transducer spacing Code Class Description 0 Success No command specific errors 1 4 Undefined 5 Error Too few data bytes received 6 Error Device specific command...

Страница 131: ...Success No command specific errors 1 Undefined 2 Error Invalid selection 3 4 Undefined 5 Error Too few data bytes received 6 Error Device specific command error 7 Error In write protect mode 8 15 Unde...

Страница 132: ...ad Reynolds KFactor table Table 144 Response data bytes for read Reynolds KFactor table Byte Format Description 0 Unsigned 8 Velocity KFactor index 1 6 Byte Format Description 0 Unsigned 8 Velocity KF...

Страница 133: ...Undefined 2 Error Invalid selection 3 4 Undefined 5 Error Too few data bytes received 6 Error Device specific command error 7 127 Undefined Byte Format Description 0 Unsigned 8 Reynolds correction 0...

Страница 134: ...few data bytes received 6 Error Device specific command error 7 Error In write protect mode 8 15 Undefined 16 Error Access restricted 17 127 Undefined Byte Format Description 0 Unsigned 8 Velocity KF...

Страница 135: ...eynolds KFactor Table Byte Format Description 0 Unsigned 8 Reynolds KFactor index 1 6 1 4 Float Reynolds value 5 8 Float Reynolds KV value Byte Format Description 0 Unsigned 8 Reynolds KFactor index 1...

Страница 136: ...ion limit 3 Velocity low limit 4 Velocity high limit 5 Amp disc min 6 Amp disc max 7 Signal low limit 8 Sound speed limit 9 Errors allowed Byte Format Description 0 Unsigned 8 Error limit 1 Correlatio...

Страница 137: ...ad flowmeter S N Byte Format Description 0 Unsigned 8 Signal setup type 1 Delta T offset 2 Percentage peak 3 Min peak percentage 4 Max peak percentage Byte Format Description 0 Unsigned 8 Signal setup...

Страница 138: ...ormat Description 0 Unsigned 8 Signal setup type 1 Electronic S N 2 UP sensor 3 S N 4 DN sensor S N 1 4 Unsigned 8 S N Code Class Description 0 Success No command specific errors 1 Undefined 2 Error I...

Страница 139: ...mand error 7 127 Undefined Byte Format Description 0 Unsigned 8 Error limit Correlation peak limit Acceleration limit Velocity low limit Velocity high limit Amp disc min Amp disc max Signal low limit...

Страница 140: ...Undefined 16 Error Access restricted 17 127 Undefined Byte Format Description 0 Unsigned 8 Signal setup type Delta T offset percentage peak Min peak percentage Max peak percentage 1 4 Float Signal set...

Страница 141: ...igned 8 Reset type 1 Reset error log 2 Forward inventory 3 Reverse inventory 4 Net inventory 5 Inventory time 6 All 7 Inventory Byte Format Description 0 Unsigned 8 Reset type Reset error log Forward...

Страница 142: ...ad the factory setting Table 178 Command specific response codes for read the factory setting Byte Format Description None Byte Format Description 0 Unsigned 8 Response time 0 5s 1s 5s 10s 30s 60s 1 4...

Страница 143: ...write the factory setting Table 181 Command specific response codes for write the factory setting Byte Format Description 0 Unsigned 8 Response time 0 5s 1s 5s 10s 30s 60s 1 4 Unsigned 32 Sample size...

Страница 144: ...odes for reset to factory setting 5 Error Too few data bytes received 6 Error Device specific command error 7 Error In write protect mode 8 15 Undefined 16 Error Access restricted 17 127 Undefined Byt...

Страница 145: ...Low signal Error 4 7 2 Sound speed error Error 4 7 3 Velocity range Error 4 7 4 Signal quality Error 4 7 5 Cycle skip Error 4 7 6 Reserve 7 Reserve 1 0 Reserve 1 Reserve 2 Reserve 3 Reserve 4 Reserve...

Страница 146: ...9 68 Volumetric Inventory totalizer time 10 70 Time Mass flow 11 72 Mass flow Sound speed 12 67 Velocity Reynolds 13 0 Not classified Kfactor 14 0 Not classified Transit time up 15 70 Time Transit ti...

Страница 147: ...et per minute 186 Standard cubic feet per second 29 Cubic meter per day 19 Cubic meter per hour 131 Cubic meters per minute 28 Cubic meters per second 240 Million cubic meters per day 187 Standard cub...

Страница 148: ...ric ton 168 Standard cubic feet 63 Pound 247 Kilo pound 64 Short tons 69 Length 44 Feet 47 Inch 45 Meter 49 Millimeter 70 Time 172 Nanoseconds 171 Microseconds 170 Milliseconds 51 Seconds 50 Minute 52...

Страница 149: ...our 86 Short ton per day 73 Mass per volume 94 Pounds per cubic feet 92 Kilograms per cubic meter 74 Viscosity 54 Centistokes 248 Square meter per sec 81 Analytical 57 Percent 96 Acceleration 171 Feet...

Страница 150: ...Chapter 6 Communication 140 AquaTrans AT600 User s manual no content intended for this page...

Страница 151: ...ading in application for 2 in 50 mm pipe and 1 ft s 0 3 m s velocity 2 of reading in application for 2 in 50 mm pipe and 1 ft s 0 3 m s velocity 0 5 in field calibration Note Installation assumes a fu...

Страница 152: ...60 Hz Optional 12 to 28 VDC 5 Power consumption In rush 10 W Normal operation 5 W Operating temperature 4 to 131 F 20 to 55 C Storage temperature 40 to 158 F 40 to 70 C Outputs based on configuration...

Страница 153: ...STM A316 Fixture body stainless steel Note Contact Panametrics for other transducer models Temperature range AT6 transducer 40 to 302 F 40 to 150 C C RS transducer 40 to 302 F 40 to 150 C UTX transduc...

Страница 154: ...26 AWG according to UL CUL range 14 to 28 A 4 2 Cable fixing requirement and gland torque Refer to Figure 24 on page 20 for the gland hole position To make a reliable IP67 sealing performance of the...

Страница 155: ...details of the service should be recorded in this appendix An accurate service history of the meter can prove very helpful in troubleshooting any future problems B 2 Data entry Record the complete an...

Страница 156: ...peration should be entered below Table 189 Initial settings Parameter Initial value Pipe OD Pipe ID Pipe wall thickness Pipe material Pipe sound speed Lining thickness Lining material Transducer ID Tr...

Страница 157: ...he system Table 190 Diagnostic parameters Parameter Initial value Velocity Actual volumetric Standardized volumetric Fwd batch totals Rev batch totals Net batch totals Batch totalizer time Fwd invento...

Страница 158: ...Appendix B Specifications 148 AquaTrans AT600 User s Manual no content intended for this page...

Страница 159: ...update make sure that the AT600 is in normal measurement mode The AT600 main power must remain ON throughout the firmware update DO NOT turn the main power OFF until the firmware update has been compl...

Страница 160: ...ile see figure 37 below Installation of the software on the PC is not required Figure 37 Running the update exe software 4 Click on the COM port button and enter the specific com port on the PC that i...

Страница 161: ...n the clear S2 warning button see bottom red box in figure 39 below Figure 39 Clearing an S2 warning 3 After about 15 seconds a screen similar to figure 40 below is displayed Click OK and then reboot...

Страница 162: ...Appendix C Specifications 152 AquaTrans AT600 User s Manual no content intended for this page...

Страница 163: ...AquaTrans AT600 User s Manual 153 Appendix D Menu maps Appendix D Menu maps D 1 The display measurement menu Figure 41 The display measurement menu...

Страница 164: ...Appendix D Menu maps 154 AquaTrans AT600 User s Manual D 2 The main menu Figure 42 The main menu...

Страница 165: ...AquaTrans AT600 User s Manual 155 Appendix D Menu maps D 3 The main menu user preferences menu Figure 43 The main menu user preferences menu...

Страница 166: ...Appendix D Menu maps 156 AquaTrans AT600 User s Manual D 4 The main menu inputs outputs menu Figure 44 The main menu inputs outputs menu...

Страница 167: ...AquaTrans AT600 User s Manual 157 Appendix D Menu maps D 5 The main menu sensor setup menu Figure 45 The main menu sensor setup menu...

Страница 168: ...Appendix D Menu maps 158 AquaTrans AT600 User s Manual D 6 The main menu calibration menu Figure 46 The main menu calibration menu...

Страница 169: ...AquaTrans AT600 User s Manual 159 Appendix D Menu maps D 7 The main menu advanced menu Figure 47 The main menu advanced menu...

Страница 170: ...Appendix D Menu maps 160 AquaTrans AT600 User s Manual D 8 The main menu factory menu Figure 48 The main menu factory menu...

Страница 171: ...lar purpose and warranties arising from course of dealing or usage or trade Return Policy If a Panametrics Sensing instrument malfunctions within the warranty period the following procedure must be co...

Страница 172: ...Warranty 162 AquaTrans AT600 User s Manual...

Страница 173: ......

Страница 174: ...respective holders BH001C11 EN F Sep 2023 Scan here for Panametrics support and service information Customer Support Centers U S A The Boston Center 1100 Technology Park Drive Billerica MA 01821 U S...