AquaTrans™ AT600 User’s manual
79
Chapter 6. Communication
Chapter 6. Communication
6.1
MODBUS
6.1.1
Introduction
The AT600 flow meter generally follows the standard MODBUS communications protocol defined by the MODBUS
APPLICATION PROTOCOL SPECIFICATION V1.1b, which is available at www.modbus.org. With this reference as a guide, an
operator can use any MODBUS master to communicate with the AT600 flow meter.
For the AT600, there are two limits of this implementation:
1.
The AT600 supports only four of the standard function codes. These are read holding registers (0x03), read input
registers (0x04), write multiple registers (0x10), and read file record (0x14).
2.
The AT600 needs a 15 ms gap between Modbus requests. The prime objective of the flow meter is to measure flow
and drive the output, so the Modbus server has a low priority.
6.1.2
MODBUS register map
Table 7: MODBUS register map
Register
(in hex)
Register
(in decimal)
Access
level
Description
RO/RW Format
100
100
256
User
Product short tag
RW
CHAR * 16
108
264
User
Product long tag
RW
CHAR * 32
118
280
User
Product message (for HART)
RW
CHAR * 32
128
296
User
Product descriptor (for HART)
RW
CHAR * 16
140
140
320
User
Product electronic serial number
RW
CHAR * 16
148
328
User
Product fixture serial number
RW
CHAR * 16
150
336
User
Product transducer1 serial number
RW
CHAR * 16
158
344
User
Product transducer 2 serial number
RW
CHAR * 16
300
300
768
RO
Main hardware version
RO
CHAR * 8
304
772
RO
Option hardware version
RO
CHAR * 8
308
772
RO
Main software version
RO
CHAR * 8
500
500
1280
User
Global unit group 1 for actual volumetric
RW
INT32
502
1282
User
Global unit group 2 for day
RW
INT32
504
1284
User
Global unit group 3 for dB
RW
INT32
506
1286
User
Global unit group 4 for density
RW
INT32
508
1288
User
Global unit group 5 for dimension
RW
INT32
50A
1290
User
Global unit group 6 for Hz
RW
INT32
50C
1292
User
Global unit group 7 for viscosity
RW
INT32
50E
1294
User
Global unit group 8 for mA
RW
INT32
510
1296
User
Global unit group 9 for mass
RW
INT32
512
1298
User
Global unit group 10 for milli second
RW
INT32
514
1300
User
Global unit group 11 for nano second
RW
INT32
516
1302
User
Global unit group 12 for percent
RW
INT32
518
1304
User
Global unit group 13 for second
RW
INT32
51A
1306
User
Global unit group 14 for standard volumetric RW
INT32
51C
1308
User
Global unit group 15 for thermo
RW
INT32
Содержание Panametrics AquaTrans AT600
Страница 1: ...panametrics com BH001C11 Rev F AquaTrans AT600 User s manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 10: ...Contents viii AquaTrans AT600 User s manual no content intended for this page...
Страница 20: ...Chapter 1 General installation instructions 10 AquaTrans AT600 User s manual no content intended for this page...
Страница 36: ...Chapter 3 Wiring the AT600 electronics 26 AquaTrans AT600 User s manual no content intended for this page...
Страница 84: ...Chapter 4 Initial setup and programming 74 AquaTrans AT600 User s manual no content intended for this page...
Страница 150: ...Chapter 6 Communication 140 AquaTrans AT600 User s manual no content intended for this page...
Страница 158: ...Appendix B Specifications 148 AquaTrans AT600 User s Manual no content intended for this page...
Страница 162: ...Appendix C Specifications 152 AquaTrans AT600 User s Manual no content intended for this page...
Страница 164: ...Appendix D Menu maps 154 AquaTrans AT600 User s Manual D 2 The main menu Figure 42 The main menu...
Страница 172: ...Warranty 162 AquaTrans AT600 User s Manual...
Страница 173: ......