AquaTrans™ AT600 User’s manual
97
Chapter 6. Communication
Command 168 (0xA8): Enter/exit fixed loop current
Enter or exit the fixed mode of loop current.
Table 50: Request data bytes for enter/exit fixed loop current
Table 51: Response data bytes for enter/exit fixed loop current
Table 52: Command-specific response code for enter/exit fixed loop currents
Command 169 (0xA9): Set loop current zero
This command is to trim the zero or lower endpoint value of the loop current to its minimum.
Table 53: Request data bytes for set loop current zero
Byte
Format
Description
0
Unsigned-8
Fixed current level:
0: Exit fixed loop current;
1: fixed 4 mA;
2: fixed 20mA;
3: fixed percentage of scale
Byte
Format
Description
0
Unsigned-8
Fixed current level:
0: Exit fixed loop current;
1: fixed 4 mA;
2: fixed 20mA;
3: fixed percentage of scale;
Code Class
Description
0
Success
No command-specific errors
1 - 4
Undefined
5
Error
Too few data bytes received
6
Error
Device-specific command error
7
Error
In write protect mode
8 - 10
Undefined
11
Error
Loop current not active
12 - 15
Undefined
16
Error
Access restricted
17 - 31
Undefined
32
Error
Busy
33 - 127
Undefined
Byte
Format
Description
0 - 3
Float
Externally measured loop current level, units of
milliamperes
Содержание Panametrics AquaTrans AT600
Страница 1: ...panametrics com BH001C11 Rev F AquaTrans AT600 User s manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 10: ...Contents viii AquaTrans AT600 User s manual no content intended for this page...
Страница 20: ...Chapter 1 General installation instructions 10 AquaTrans AT600 User s manual no content intended for this page...
Страница 36: ...Chapter 3 Wiring the AT600 electronics 26 AquaTrans AT600 User s manual no content intended for this page...
Страница 84: ...Chapter 4 Initial setup and programming 74 AquaTrans AT600 User s manual no content intended for this page...
Страница 150: ...Chapter 6 Communication 140 AquaTrans AT600 User s manual no content intended for this page...
Страница 158: ...Appendix B Specifications 148 AquaTrans AT600 User s Manual no content intended for this page...
Страница 162: ...Appendix C Specifications 152 AquaTrans AT600 User s Manual no content intended for this page...
Страница 164: ...Appendix D Menu maps 154 AquaTrans AT600 User s Manual D 2 The main menu Figure 42 The main menu...
Страница 172: ...Warranty 162 AquaTrans AT600 User s Manual...
Страница 173: ......