Chapter 1. General installation instructions
6
AquaTrans™ AT600 User’s manual
The installation base of AT600 electronics enclosure can be rotated 90° to keep a horizontal view of the user interface
in any mounting orientation. See figure 6 below for drawings of the AT600 mounting base.
Figure 6: The AT600 mounting base
1.6
Calculating the transducer spacing
Before installing the clamp-on fixture(s) and transducers, you must program the AT600 to calculate the required
transducer spacing for your planned installation. To accomplish this task, go to “Sensor setup” on page 56 and follow
the instructions in that section. After obtaining the required transducer spacing value, return here and continue to the
next section.
Horizontal
Vertical
Содержание Panametrics AquaTrans AT600
Страница 1: ...panametrics com BH001C11 Rev F AquaTrans AT600 User s manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 10: ...Contents viii AquaTrans AT600 User s manual no content intended for this page...
Страница 20: ...Chapter 1 General installation instructions 10 AquaTrans AT600 User s manual no content intended for this page...
Страница 36: ...Chapter 3 Wiring the AT600 electronics 26 AquaTrans AT600 User s manual no content intended for this page...
Страница 84: ...Chapter 4 Initial setup and programming 74 AquaTrans AT600 User s manual no content intended for this page...
Страница 150: ...Chapter 6 Communication 140 AquaTrans AT600 User s manual no content intended for this page...
Страница 158: ...Appendix B Specifications 148 AquaTrans AT600 User s Manual no content intended for this page...
Страница 162: ...Appendix C Specifications 152 AquaTrans AT600 User s Manual no content intended for this page...
Страница 164: ...Appendix D Menu maps 154 AquaTrans AT600 User s Manual D 2 The main menu Figure 42 The main menu...
Страница 172: ...Warranty 162 AquaTrans AT600 User s Manual...
Страница 173: ......