AquaTrans™ AT600 User’s manual
9
Chapter 1. General installation instructions
Table 2: AT600 clamp-on fixture installation
Figure 9: Flowchart to transducer installation instructions
Pipe size range
Transducer
frequency
(MHz)
Number of
traverses
Transducer
spacing
(mm)
Number of
fixtures
mm
inches
50 to 100
2 to 4
2
4
32 to 250
1
100 to 150
4 to 6
2
2
32 to 250
1
50 to 150
2 to 6
2
1
0 to 250
2
100 to 300
4 to 12
1
2
50 to 320
1
300 to 600
12 to 24
1
2
320 to 940
2
600 to 1500
24 to 60
1
1
>320
2
200 to 300
8 to 12
0.5
2
50 to 320
1
300 to 900
12 to 36
0.5
2
320 to 940
2
>900
>36
0.5
1
>320
2
Содержание Panametrics AquaTrans AT600
Страница 1: ...panametrics com BH001C11 Rev F AquaTrans AT600 User s manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 10: ...Contents viii AquaTrans AT600 User s manual no content intended for this page...
Страница 20: ...Chapter 1 General installation instructions 10 AquaTrans AT600 User s manual no content intended for this page...
Страница 36: ...Chapter 3 Wiring the AT600 electronics 26 AquaTrans AT600 User s manual no content intended for this page...
Страница 84: ...Chapter 4 Initial setup and programming 74 AquaTrans AT600 User s manual no content intended for this page...
Страница 150: ...Chapter 6 Communication 140 AquaTrans AT600 User s manual no content intended for this page...
Страница 158: ...Appendix B Specifications 148 AquaTrans AT600 User s Manual no content intended for this page...
Страница 162: ...Appendix C Specifications 152 AquaTrans AT600 User s Manual no content intended for this page...
Страница 164: ...Appendix D Menu maps 154 AquaTrans AT600 User s Manual D 2 The main menu Figure 42 The main menu...
Страница 172: ...Warranty 162 AquaTrans AT600 User s Manual...
Страница 173: ......