The followings table shows the preset key assignment and which combinations are possible:
Return to factory settings
In order to return to factory settings:
►
Start menu (button A
20
(up), press and hold).
►
Mark "System setting".
►
Actuate button B
21
(right). A submenu appears.
►
Select "Return to default settings".
End application
►
Press the ”Power off” switch
1
on the camera control unit.
►
Disconnect light guide cable.
►
Disconnect endoscope.
►
To completely disconnect the device from the power supply, remove the
plug from the mains connector
6
on the rear of the device.
6.
Validated reprocessing procedure
6.1
General safety instructions
Note
Adhere to national statutory regulations, national and international stan-
dards and directives, and local, clinical hygiene instructions for sterile pro-
cessing.
Note
For patients with Creutzfeldt-Jakob-disease (CJD), suspected CJD or possi-
ble variants of CJD, observe the relevant national regulations concerning
the reprocessing of products.
Note
Successful processing of this medical device can only be ensured if the pro-
cessing method is first validated. The operator/sterile processing technician
is responsible for this.
The recommended chemistry was used for validation.
Note
If there is no final sterilization, then a virucidal disinfectant must be used.
6.2
Single-use products
WARNING
Risk of infection for patients and/or users and impairment of product
functionality due to reuse. Risk of injury, illness or death due to con-
tamination and/or impaired functionality of the product!
►
Do not reprocess the product!
6.3
General information
Dried or affixed surgical residues can make cleaning more difficult or inef-
fective and lead to corrosion. Therefore the time interval between applica-
tion and processing should not exceed 6 h; also, neither fixating pre-
cleaning temperatures >45 °C nor fixating disinfecting agents (active
ingredient: aldehydes/alcohols) should be used.
Excessive measures of neutralizing agents or basic cleaners may result in
a chemical attack and/or to fading and the laser marking becoming
unreadable visually or by machine for stainless steel.
Residues containing chlorine or chlorides e.g. in surgical residues, medi-
cines, saline solutions and in the service water used for cleaning, disinfec-
tion and sterilization will cause corrosion damage (pitting, stress corro-
sion) and result in the destruction of stainless steel products. These must
be removed by rinsing thoroughly with demineralized water and then dry-
ing.
Additional drying, if necessary.
Function/push-
button
Up short
Top long
Down short
Down long
Left short
Left long
Right short
Right long
Menu
■
(default
and fixed)
White balance
■
■
■
(default)
Light source on/off
■
(default)
■
■
Photo
■
■
■
■
(default)
■
■
■
Video
■
■
■
■
■
(default)
■
■
Brig
■
■
■
■
■
■
■
Brightness -
■
■
■
■
■
■
■
Satu
■
■
■
■
■
■
■
Saturation -
■
■
■
■
■
■
■
Digital zoom +
■
■
■
■
■
■
■
Digital zoom -
■
■
■
■
■
■
■
Sha
■
■
■
■
■
■
■
Sharpness -
■
■
■
■
■
■
■
Co
■
■
■
■
■
■
■
Contrast -
■
■
■
■
■
■
■
Empty
■
(default)
■
(default)
■
■
■
■
(default)
■
Designation
Art. no.
Disposable sterile cover
JG904