11
en
8.
Maintenance
To ensure reliable operation, the product must be maintained in accor-
dance with the maintenance labeling or at least once a year.
YYYY-MM
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
9.
Troubleshooting list
►
Have defective products repaired by Aesculap Technical Service, see Technical Service.
Issue
Cause
Detection
Remedy
Product not running
No battery
No battery in the battery compartment Insert battery.
Battery not charged
No beep when inserting the battery
Charge battery in charger.
Battery defective
No signal tone when inserting the bat-
tery
Have the battery repaired by the manufacturer.
Product in secured OFF posi-
tion
Safety catch is in OFF position
Turn the safety catch to the ON position.
Product defective
Product not running
Have the product repaired by the manufacturer.
Product is getting too hot Excessive use
Product is heating up
Observe instructions for use (nominal operating
mode).
Reprocessing/maintenance
carried out incorrectly
Product heating
Follow operating instructions (preparation, care).
Preventative: lubricate product before every steril-
ization.
Fall damage, product defec-
tive
Product heating
Have the product repaired by the manufacturer.
Blunt tool
Tool and product are heating up
Change tool.
Insufficient power
Product defective
Inadequate product power
Follow operating instructions (preparation, care).
Preventive measure: Lubricate product before
every sterilization.
Have the product repaired by the manufacturer.
Large amount of heat generated after a
short time
Follow operating instructions (operating mode).
Have the manufacturer repair the product
Blunt tool
Tool blades worn out
Replace tool.
Loud running noise
Defective product drive/ball
bearings
Loud, perceptible noise during opera-
tion
Observe instructions for use (processing, care).
Preventive measure: Lubricate product before
every sterilization.
Have the manufacturer repair the product
Lid cannot be
attached/removed
Lid not compatible
Lid does not click into place
Use the compatible lid for GA331.
Lid deformed/defective
Lid is difficult or cannot be
mounted/removed
Have the manufacturer repair the lid.
Connection to product
deformed/defective
Lid is hard or impossible to fit/remove
Have the product repaired by the manufacturer.
Battery cannot be
inserted/removed
Battery not compatible
Battery cannot be completely inserted
into the battery compartment
Use the compatible battery for GA331.
Battery deformed/defective
Battery is is difficult to insert/remove
or cannot be inserted/removed
Have the manufacturer repair the battery.
Battery compartment on
product deformed/defective
Battery is difficult or cannot be
mounted/removed
Have the product repaired by the manufacturer