146
Division
Adjustment
Serial No.
Section
Adjusting the main board
Tools and jigs to be used
(1) V-2100G Simulator
(2) Digital multimeter
(3) Flat-head screwdriver (small)
No Adjustment
procedure Remarks
1
Preparation
Pull the relay box out of the main body, and get the main board ready for adjustment.
Connect the V-2100G Simulator (hereinafter referred to as Simulator) to the incubator.
Turn the incubator power switch on.
2
Checking the supply voltage
Connect the digital multimeter to the following test pins (TPs) and check each voltage (Vdc).
CPU power source: TP24 (GND) – TP25 5.00
±
0.25V
Oxygen control power source: TP24 (GND)
-
TP23 12.0
±
0.6V
TP24 (GND)
-
TP27 -12.0
±
0.6V
TP25
TP24
TP23
TP27
Содержание V-2200
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 14: ...12 4 2 Display Control Panel...
Страница 186: ...184 8 Control Unit Block Diagram This block diagram is not available for units destined for the US...
Страница 187: ...185 Available only for units destined for the US...
Страница 188: ...186 9 Wiring Diagram 9 1 Main Unit C60F0243 Available only for units with serial No 1679999 or smaller...
Страница 189: ...187 C60F0263 Available only for units with serial No 1680000 or higher...
Страница 190: ...188 C60F1003 Available only for units destined for the US...
Страница 191: ...189 9 2 HL stand...
Страница 192: ...190...
Страница 193: ...191 10 Replacement Parts List Replacement Parts List...
Страница 194: ...192 10 1 Pole Arm Upper Part of the Main Body Accessories and Humidity Chamber Section Fig 1...
Страница 196: ...194 10 2 Hood Section Fig 2...
Страница 198: ...196 10 3 Main Body Fig 3...
Страница 202: ...200 10 5 HL Stand Fig 5...
Страница 211: ......
Страница 212: ...C60S220 201109...