159
Division
Adjustment
Serial No.
Section
Resetting the record
Tools and jigs to be used
(1) Flat-head screwdriver (small)
No Adjustment
procedure
Remarks
1
Preparation
Disassemble the display and expose the main board. Then, turn the dip
switches SW4-1 and SW4-2 ON.
SW4-1: To clear the past data
SW4-2: To clear the parts replacement records
2
Reassemble the display and connect it to the incubator. Then, turn the power
switch on and wait for 100 seconds.
3
Disassemble the display again, and turn the dip switches SW4-1 and SW4-2
OFF
4
Reassemble the display.
SW4
②