152
Division
Adjustment
Serial No.
Section
Setting the dip switch of the main board
Tools and jigs to be used
(1) Flat-head screwdriver (small)
※
C500723, C500724
No Adjustment
procedure
Remarks
2
Each volume and tone call be changed using the dip switch (SW3) on the main board.
If desired, change the settings as shown above.
SW
3
1
2
3
4
5
6
7
8
Factory default setting
ON
OFF
ON
ON
ON
OFF
ON
ON
①
deviation alarm, high temperature alarm, skin temperature probe alarm, fan alarm, system failure alar, and humidity sensor
alarm
1:
Tone
(
ON
:
high
OFF
:
low
)
2:
Volume
(
ON
:
small
OFF
:
loud
)
②
pulse synchronous sound
3:
V
olume
(
ON
:
small
OFF
:
loud
)
③
oxygen concentration alarm, oxygen sensor alarm, oxygen flow rate alarm, oxygen sensor position error
alarm
4:
Volume
(
ON
:
small
OFF
:
loud
)
5:
Tone
(
ON
:
high
OFF
:
low
)
④
SpO
2
high/low limit alarms, pulse rate high/low limit alarms
6:
Tone
(
ON
:
high
OFF
:
low
)
7:
Volume
(
ON
:
small
OFF
:
lound
)
8:
Not used
Содержание V-2200
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 14: ...12 4 2 Display Control Panel...
Страница 186: ...184 8 Control Unit Block Diagram This block diagram is not available for units destined for the US...
Страница 187: ...185 Available only for units destined for the US...
Страница 188: ...186 9 Wiring Diagram 9 1 Main Unit C60F0243 Available only for units with serial No 1679999 or smaller...
Страница 189: ...187 C60F0263 Available only for units with serial No 1680000 or higher...
Страница 190: ...188 C60F1003 Available only for units destined for the US...
Страница 191: ...189 9 2 HL stand...
Страница 192: ...190...
Страница 193: ...191 10 Replacement Parts List Replacement Parts List...
Страница 194: ...192 10 1 Pole Arm Upper Part of the Main Body Accessories and Humidity Chamber Section Fig 1...
Страница 196: ...194 10 2 Hood Section Fig 2...
Страница 198: ...196 10 3 Main Body Fig 3...
Страница 202: ...200 10 5 HL Stand Fig 5...
Страница 211: ......
Страница 212: ...C60S220 201109...