155
Division
Adjustment
Serial No.
Section
Adjusting the DC motor power supply board
Tools and jigs to be used
(1) Non-contact type tachometer (2) Flat-head screwdriver (small)
No Adjustment
procedure
Remarks
1
Preparation
Open the hood of the incubator and remove the mattress platform and the top
board.
With the casing placed on, affix the reflector seal supplied with the non-
contact type tachometer to the fan.
Remove the incubator panel, and get the DC motor power supply board ready
for adjustment.
Turn the incubator power switch on.
2
Adjusting the revolution rate of the fan
Direct the non-contact type tachmeter toward the reflector seal affixed to the
fan, and measure the revolution rate of the fan.
Adjust VR1 on the DC motor powr supply board for the revolution rate to be
1250
±
30rpm.
VR1
Содержание V-2200
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 14: ...12 4 2 Display Control Panel...
Страница 186: ...184 8 Control Unit Block Diagram This block diagram is not available for units destined for the US...
Страница 187: ...185 Available only for units destined for the US...
Страница 188: ...186 9 Wiring Diagram 9 1 Main Unit C60F0243 Available only for units with serial No 1679999 or smaller...
Страница 189: ...187 C60F0263 Available only for units with serial No 1680000 or higher...
Страница 190: ...188 C60F1003 Available only for units destined for the US...
Страница 191: ...189 9 2 HL stand...
Страница 192: ...190...
Страница 193: ...191 10 Replacement Parts List Replacement Parts List...
Страница 194: ...192 10 1 Pole Arm Upper Part of the Main Body Accessories and Humidity Chamber Section Fig 1...
Страница 196: ...194 10 2 Hood Section Fig 2...
Страница 198: ...196 10 3 Main Body Fig 3...
Страница 202: ...200 10 5 HL Stand Fig 5...
Страница 211: ......
Страница 212: ...C60S220 201109...