33
5-3-2. Mattress Platform and Parts Beneath
Hold the grip on the front of the hood and raise the
hood. The hood stopper on the right side of the
hood will be folded fully backward, and the hood
will be locked in place.
Clean and disinfect the heater and its
surroundings only when the heater has
cooled down sufficiently, or you may get
burned.
To raise the hood, be sure to hold the grip.
If you hold the hood by any other part, the
hood will slip and cause a hazard.
Be sure to check that the hood stopper has
been folded fully backward. If it has been
folded only insufficiently, the hood will fall
forward and cause a hazard.
To raise or lower the hood, be sure to lower
the mattress platform first. If the hood is
raised or lowered with the mattress platform
tilted or raised, it will hit the mattress
platform and be damaged.
When you raise the hood, be very careful
not to let it hit the display/control panel so
that it may not get broken.
Before you raise the hood, pull out the
weight monitor cable connected to the main
body of the incubator. If you raise the hood
without pulling out the weight monitor cable,
the cable will be pulled by force and a wire
inside or the cable itself may get broken.
●
Mattress Platform
Hold the mattress platform by both edges and push
it fully backward.
When it is removed from the rails, lift it up and take
it out. Wipe it clean with a soft cloth dampened
with a disinfectant solution. To replace it to its
former position, place it fully backward and then
pull it fully forward to make sure that it is in
engagement with the rails.
Содержание V-2200
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 14: ...12 4 2 Display Control Panel...
Страница 186: ...184 8 Control Unit Block Diagram This block diagram is not available for units destined for the US...
Страница 187: ...185 Available only for units destined for the US...
Страница 188: ...186 9 Wiring Diagram 9 1 Main Unit C60F0243 Available only for units with serial No 1679999 or smaller...
Страница 189: ...187 C60F0263 Available only for units with serial No 1680000 or higher...
Страница 190: ...188 C60F1003 Available only for units destined for the US...
Страница 191: ...189 9 2 HL stand...
Страница 192: ...190...
Страница 193: ...191 10 Replacement Parts List Replacement Parts List...
Страница 194: ...192 10 1 Pole Arm Upper Part of the Main Body Accessories and Humidity Chamber Section Fig 1...
Страница 196: ...194 10 2 Hood Section Fig 2...
Страница 198: ...196 10 3 Main Body Fig 3...
Страница 202: ...200 10 5 HL Stand Fig 5...
Страница 211: ......
Страница 212: ...C60S220 201109...