62
occurring at the same time, not the message but the
alarm mark will appear.
Action to take
: Check the SpO
2
sensor, the patient cable and the connector for any broken wire or
defective contact.
Alarm reset
: The alarm will stop if the SpO
2
sensor is connected properly as it should. If the
alarm silence/reset switch is pressed, the alarm will be cancelled and the SpO
2
function will be stopped.
6-6-11. Oxygen Sensor Alarm
If the oxygen sensor has not been calibrated properly,
an oxygen sensor alarm message will appear and an
audible alarm (“pip pip peep pip pip”) will sound. At
this time the alarm lamp on the display/control panel
will flash. If another alarm should be occurring at
the same time, not the message but the alarm mark
will appear.
Action to take
: Recalibrate the sensor. If the oxygen sensor cannot be calibrated properly in
recalibration, it should be regarded as faulty.
Alarm reset
: The alarm will stop if the oxygen sensor is calibrated properly in recalibration.
6-6-12. Oxygen Concentration Alarm
If the oxygen concentration in the incubator deviates
from the selected oxygen concentration by ±3% or
more, an oxygen concentration alarm message will
appear and an audible alarm (“pip pip peep pip pip”)
will sound. At this time the alarm lamp on the
display/control panel will flash. If another alarm
should be occurring at the same time, not the message
but the alarm mark will appear.
Action to take
: Check that the access ports and the front admittance panel are closed and that
oxygen is sufficiently supplied.
Alarm reset
: The audible alarm can be silenced for two minutes by pressing the alarm
silence/reset switch. The alarm will sound again after two minutes unless the
normal condition has returned.
The unit is so designed that the oxygen concentration alarm does not activate for 40
minutes after the oxygen controller is turned on or after the oxygen concentration is
changed to another level. However, the oxygen concentration alarm will begin to
activate even within 40 minutes of the above-mentioned events if the oxygen
concentration should reach the oxygen concentration alarm range (selected value
±
2%). If the oxygen concentration level is set to 23% or lower, the oxygen
concentration alarm will not function. In this case, monitor the oxygen
concentration continuously and closely.
Содержание V-2200
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 14: ...12 4 2 Display Control Panel...
Страница 186: ...184 8 Control Unit Block Diagram This block diagram is not available for units destined for the US...
Страница 187: ...185 Available only for units destined for the US...
Страница 188: ...186 9 Wiring Diagram 9 1 Main Unit C60F0243 Available only for units with serial No 1679999 or smaller...
Страница 189: ...187 C60F0263 Available only for units with serial No 1680000 or higher...
Страница 190: ...188 C60F1003 Available only for units destined for the US...
Страница 191: ...189 9 2 HL stand...
Страница 192: ...190...
Страница 193: ...191 10 Replacement Parts List Replacement Parts List...
Страница 194: ...192 10 1 Pole Arm Upper Part of the Main Body Accessories and Humidity Chamber Section Fig 1...
Страница 196: ...194 10 2 Hood Section Fig 2...
Страница 198: ...196 10 3 Main Body Fig 3...
Страница 202: ...200 10 5 HL Stand Fig 5...
Страница 211: ......
Страница 212: ...C60S220 201109...