179
Major division
Product inspection
Serial No.
Subdivision
Leakage current test
Tools and jigs to be used
(1) Leakage current tester
No Inspection
procedure Remarks
1
Connect the leakage current tester to the incubator.
2
Leakage current
Turn the incubator power switch on and set the incubator as follows.
Heater output: Maximum Humidification heater: ON Oxygen controller: ON
Measure the following leakage currents. Check that the measurements do not
exceed the following values.
Each leakage current:
≦
0.5mA
Enclosure leakage current:
≦
0.1mA (Accessible conductive part: the hinges of
the hood)
Patient leakage current 1:
≦
0.1mA (Measure the skin temperature probe and
the SpO
2
sensor at a time.)
Содержание V-2200
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 14: ...12 4 2 Display Control Panel...
Страница 186: ...184 8 Control Unit Block Diagram This block diagram is not available for units destined for the US...
Страница 187: ...185 Available only for units destined for the US...
Страница 188: ...186 9 Wiring Diagram 9 1 Main Unit C60F0243 Available only for units with serial No 1679999 or smaller...
Страница 189: ...187 C60F0263 Available only for units with serial No 1680000 or higher...
Страница 190: ...188 C60F1003 Available only for units destined for the US...
Страница 191: ...189 9 2 HL stand...
Страница 192: ...190...
Страница 193: ...191 10 Replacement Parts List Replacement Parts List...
Страница 194: ...192 10 1 Pole Arm Upper Part of the Main Body Accessories and Humidity Chamber Section Fig 1...
Страница 196: ...194 10 2 Hood Section Fig 2...
Страница 198: ...196 10 3 Main Body Fig 3...
Страница 202: ...200 10 5 HL Stand Fig 5...
Страница 211: ......
Страница 212: ...C60S220 201109...