178
Major division
Product inspection
Serial No.
Subdivision
Weight monitor
Tools and jigs to be used
(1) Standard weight
(
1kg,3kg, and 5kg
)
No Inspection
procedure Remarks
6
Clearing the memory of the incubator
Turn the incubator power switch off, and then turn it on again.
Press the Screen Changeover switch to have the Trend 2 screen displayed,
and check that the plot still remains on the weight graph.
Press the Screen Changeover switch, and check that the weight display on the
numerical value screen shows “MANUAL WEIGHT MEASUREMENT ---g.”
Press and hold the Weight switch until a “pip, peep” is heard. (The internal
memory is cleared.)
Press the Screen Changeover switch to have the weight graph on the Trend 2
screen displayed again.
Check that no plot remains on the graph.
Содержание V-2200
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 14: ...12 4 2 Display Control Panel...
Страница 186: ...184 8 Control Unit Block Diagram This block diagram is not available for units destined for the US...
Страница 187: ...185 Available only for units destined for the US...
Страница 188: ...186 9 Wiring Diagram 9 1 Main Unit C60F0243 Available only for units with serial No 1679999 or smaller...
Страница 189: ...187 C60F0263 Available only for units with serial No 1680000 or higher...
Страница 190: ...188 C60F1003 Available only for units destined for the US...
Страница 191: ...189 9 2 HL stand...
Страница 192: ...190...
Страница 193: ...191 10 Replacement Parts List Replacement Parts List...
Страница 194: ...192 10 1 Pole Arm Upper Part of the Main Body Accessories and Humidity Chamber Section Fig 1...
Страница 196: ...194 10 2 Hood Section Fig 2...
Страница 198: ...196 10 3 Main Body Fig 3...
Страница 202: ...200 10 5 HL Stand Fig 5...
Страница 211: ......
Страница 212: ...C60S220 201109...