CFS300 SERIES OPERATION MANUAL
SECTION 6: FRONT PANEL OPERATION
Entire Contents Copyright
2016 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
APS CFS300 Series Power Source Operation Manual
Page 73 of 156
6.10.8
Loop Cycle Settings
Press the “
Next
” soft key to advance to the next parameter, the Loop Cycle setting if not
already there.
The Loop Cycle count indicates how many times a program will be repeated when executed.
Default is OFF (1) meaning a test program will run one time and not repeat. When set to a
value from 2 through 9999, the program will run the number of times set. A value of zero (0)
means the program will cycle continuously until stopped by the operator. Thus for a given
setting <n>:
•
1 =
Program will run once
•
0 =
Program will run continuously until manually stopped by operator
•
2-9999 = Program will run <n> times
Use the numeric keypad to enter in the Loop Cycle setting. Press the “
Enter
” soft key to
accept the parameter. To cancel editing, press the “
Esc
” soft key.
6.10.9
Results Display Mode Settings
Press the “
Next
” soft key to advance to the next parameter, the
Results
setting.
Use the “
Change
” soft key to select from the following available Results display mode
settings:
•
ALL
Result screen will display all available settings and measurements at the end
of a test.
•
P/F
Results screen will only display a full screen size PASS or FAIL banner at the
end of a test. The operator needs to acknowledge the result by pressing the
“
Exit
” soft key in order to proceed.