CFS300 SERIES OPERATION MANUAL
SECTION 10: CALIBRATION
Entire Contents Copyright
2016 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
APS CFS300 Series Power Source Operation Manual
Page 151 of 156 PN 160952-10
Enter the current reading from the ammeter with the numeric keypad. When the value has
been selected, press the “
Enter
” soft key and you will be moved to the first calibration
parameter Volt 150.0V. The soft keys “
Volt+
” and “
Volt-
” are available, if needed, to adjust
the voltage output of the instrument to set a proper peak current value if a non-
recommended load is used. If you press the “
Esc
” soft key, you will be returned to the
calibration mode screen
10.3.7
Exiting Calibration Mode
Once all calibrations are done, cycle power to the power source using the front panel
On/Off switch so the unit returns to normal operation mode.