CFS300 SERIES OPERATION MANUAL
SECTION 10: CALIBRATION
Entire Contents Copyright
2016 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
APS CFS300 Series Power Source Operation Manual
Page 149 of 156 PN 160952-10
Enter the voltage reading from the voltmeter with the numeric keypad. When the value has
been selected, press the “
Enter
” soft key and you will be moved to the next calibration
parameter Voltage 300.0V. If you press the “
Esc
” soft key, you will be returned to the
calibration mode screen.
10.3.4
Voltage: 300.0V Calibration
Use the “
∧
,
∨
" soft keys to navigate to the Voltage 300.0V parameter and press the “
Select
”
soft key.
Follow the prompt message provided on the display, and press the Test/Reset button to
move into the calibration screen for voltage. If you press the Exit soft key at this screen you
return to the calibration mode screen.
Enter the voltage reading from the voltmeter with the numeric keypad. When the value has
been selected press the Enter soft key and you will be moved to the next calibration
parameter Current xx.xA. If you press the Esc soft key you will be returned to the calibration
mode screen