3-6 Application Interface – Setup Menu Tab
Using the 2300-580-R Software with VectorStar ME7838D4
3-18
PN: 10410-00327 Rev. D
3659 0.8 mm Cal/Ver Kit & 2300-580-R PVS UG
11.
Click
Verify Files on USB Drive
. The application checks the files on the USB drive to verify that all files are
present and that each has the proper number of data points.
12.
When all information is entered and you are ready to start, click the
Setup Calibration of VNA
button. A
confirmation dialog box appears.
13.
Click
Yes
to start. Click
No
to stop the calibration
14.
If Port1/Port2 is requested, the program execution continues in either
Calibration – Port 1/Port 2” on page 3-19
or
Section 3-8, “High-Band Calibration – Port 1/Port 2”
, depending on the previous band selection.
15.
If Port3/Port4 is requested program execution continues in either
Section 3-16, “Low-Band Calibration –
Section 3-17, “High-Band Calibration – Port 3/Port 4” on page 3-42
,
depending on the previous band selection.
Figure 3-26.
ME7838D4 Application Interface – Setup Menu Tab