Using the 2300-580-R Software with VectorStar ME7838D4
3-7 Low-Band Calibration – Port 1/Port 2
3659 0.8 mm Cal/Ver Kit & 2300-580-R PVS UG
PN: 10410-00327 Rev. D
3-21
5.
The VNA measures the open/short device data, then the
VNA Calibration: Step 4 of 5
dialog box appears.
a.
Connect the Male Short #1 (1.20 mm) to VNA Port 1.
b.
Connect the Female Open to VNA Port 2.
c.
When ready to proceed, click
OK
.
6.
The VNA measures the short/open device data, then the
VNA Calibration: Step 5 of 5
dialog box appears.
a.
Remove the Short and the Open from Port 1 and Port 2, respectively.
b.
Slide the two T-R modules towards each other and connect the VNA Port 1 (Female) to VNA Port 2
(Male) directly.
c.
When ready to proceed, click
OK
.
7.
The program execution continues with
Section 3-9 “Application Interface – Main Menu Tab” on
.
Figure 3-31.
VNA Calibration: Step 4 of 5 Dialog
Figure 3-32.
VNA Calibration: Step 5 of 5 Dialog