Using the 2300-580-R Software with VectorStar ME7838D
2-5 Running the Verification Software
3659 0.8 mm Cal/Ver Kit & 2300-580-R PVS UG
PN: 10410-00327 Rev. D
2-5
2-5
Running the Verification Software
After the VNA and PC have been configured as described above, you are ready to run the program and perform
the calibration/verification. Make sure you have the following information available:
•
Serial number of the 3659 0.8 mm Calibration/Verification Kit
•
Serial number of the 3739C Broadband/Millimeter Wave Test Set
•
Model number and serial number of the Port 1 BB/mm Module
•
Model number and serial number of the Port 2 BB/mm Module
Procedure
1.
With the equipment and software configured as described in
“Configuring the Hardware” on page 2-1
,
turn on the computer and allow it to start Windows.
2.
Locate the
Anritsu 0.8 mm Verification
icon (shown below) on the desktop and double-click it.
3.
The program displays an
About
dialog box with version information. Click
OK
to continue.
4.
The
Test Operator Name
dialog box appears. Enter a user name or other identification.
5.
Click
OK
to continue. The program searches for the GPIB Board in the PC Controller that is set to GPIB
Address 0 and displays a confirmation dialog.
6.
Click
Yes
to continue. Click
No
to search for another GPIB Board.
Figure 2-4.
About Dialog
Figure 2-5.
Name Dialog