3-17 High-Band Calibration – Port 3/Port 4
Using the 2300-580-R Software with VectorStar ME7838D4
3-44
PN: 10410-00327 Rev. D
3659 0.8 mm Cal/Ver Kit & 2300-580-R PVS UG
c.
When ready to proceed, click
OK
.
6.
The
VNA Calibration: Step 5 of 5
dialog box appears.
a.
Remove the offset Shorts from Port 3 and Port 4.
b.
Slide the two T-R modules towards each other and connect the VNA Port 1 (Female) to VNA Port 2
(Male) directly.
c.
When ready to proceed, click
OK
.
7.
The program execution continues to
Section 3-18 “Application Interface – Main Menu Tab” on page 3-45
Figure 3-63.
VNA Calibration: Step 4 of 5 Dialog
Figure 3-64.
VNA Calibration: Step 5 of 5 Dialog