3-20 Matched Thru Tests – Port 3/Port 4
Using the 2300-580-R Software with VectorStar ME7838D4
3-52
PN: 10410-00327 Rev. D
3659 0.8 mm Cal/Ver Kit & 2300-580-R PVS UG
8.
When the test completes, the next action depends on what the user selected in the
Main Menu
tab. The
VNA main display should be similar to the one in
. This is a typical display immediately after
measuring the Matched Thru in the Low Band. Please note this is for reference only and that the actual
data on the VNA display may not be identical
9.
If only the
Matched Thru
button was selected, the program returns to the
. The dialog box is shown in
.
10.
If the
Run All Tests
button was selected, the program execution continues with
.
Figure 3-72.
MS4640A/B Series VNA, Typical Matched Thru Data in Low Band
Upper-left quadrant = S43 – Forward Transmission
Upper-right quadrant = S34 – Reverse Transmission
Lower-left quadrant = S33 – Forward Reflection
Lower-right quadrant = S44 – Reverse Reflection