Using the 2300-580-R Software with VectorStar ME7838G
4-12 Mismatched Thru Tests
3659 0.8 mm Cal/Ver Kit & 2300-580-R PVS UG
PN: 10410-00327 Rev. D
4-33
If the
Matched Thru
test button was selected, only the steps in this section are executed. When this test is
complete, the program returns to the main menu described
Section 4-9 “Application Interface – Main Menu
above.
The entry point for this portion of the verification software depends on the user selections made in the
Application Interface – Main Menu
Section 4-9 “Application Interface – Main Menu Tab” on
above.
Figure 4-36, “ME7838G Application Interface – Main Menu Tab” on page 4-25
above.
If only the
Mismatched Thru
button was selected above, program execution starts here.
If the
Run All Tests
button was selected above, program execution continues here after the
Matched Thru
tests
(described above in
Section 4-11 “Matched Thru Tests” on page 4-30
) have been completed.
Procedure
1.
The
Enter Serial Number Offset Set Mismatch Device
dialog box appears. Enter the serial number of the
offset mismatch device.
2.
Ensure that the correct device is connected.
•
The
Matched Thru
verification standard is the device with one scribe line.
•
The
Mismatched Thru
verification standard is the device with two scribe lines.
Note
The verification devices must be connected to the
0.8 mm Connector Coupler Ports
and not to the
VNA Ports
. Use the torque wrench supplied with the verification kit when tightening the device
connections.
Figure 4-44.
Serial Number Dialog
Note
The verification standards must be connected to the
0.8 mm Coupler Ports
and not to the
VNA
ports
.