3-20 Matched Thru Tests – Port 3/Port 4
Using the 2300-580-R Software with VectorStar ME7838D4
3-50
PN: 10410-00327 Rev. D
3659 0.8 mm Cal/Ver Kit & 2300-580-R PVS UG
3-20 Matched Thru Tests – Port 3/Port 4
If the
Run All Tests
button on the
Application Interface - Main Menu
tab was selected, the
Matched Thru
tests (this
section) are executed, followed by the
Mismatched Thru
tests described in
Section 3-21 “Mismatched Thru Tests
If the
Matched Thru
test button was selected, only the steps in this section are executed. When this test is
complete, the program returns to the main menu described
Section 3-18 “Application Interface – Main Menu
Procedure
1.
The
Enter Serial Number Match Thru
dialog box appears.
2.
Ensure that the correct device is connected.
•
The
Matched Thru
verification standard is the device with one scribe line.
•
The
Mismatched Thru
verification standard is the device with two scribe lines.
3.
Enter the serial number of the Matched Thru.
Note
The verification devices must be connected to the
0.8 mm Connector Coupler Ports
and not to the
VNA Ports
. Use the torque wrench supplied with the verification kit when tightening the device
connections.
Figure 3-69.
Enter Serial Number Thru Dialog
Note
The verification standards must be connected to the
0.8 mm Coupler Ports
and not to the
VNA
ports
.