Using the 2300-580-R Software with VectorStar ME7838D4
3-17 High-Band Calibration – Port 3/Port 4
3659 0.8 mm Cal/Ver Kit & 2300-580-R PVS UG
PN: 10410-00327 Rev. D
3-43
3.
The
VNA Calibration: Step 2 of 5
dialog box appears.
a.
Connect the Male 1.20 mm Short #1 to VNA Port 3.
b.
Connect the Female 1.63 mm Short #2 to VNA Port 4.
c.
When ready to proceed, click
OK
.
4.
The
VNA Calibration: Step 3 of 5
dialog box appears.
a.
Connect the Male 1.63 mm Short #2 to VNA Port 3.
b.
Connect the Female 2.06 mm Short #3 to VNA Port 4.
c.
When ready to proceed, click
OK
.
5.
The
VNA Calibration: Step 4 of 5
dialog box appears.
a.
Connect the Male 2.06 mm Short #3 to VNA Port 3.
b.
Connect the Female 1.20 mm Short #1 to VNA Port 4.
Figure 3-61.
VNA Calibration: Step 2 of 5 Dialog
Figure 3-62.
VNA Calibration: Step 3 of 5 Dialog