3-17 High-Band Calibration – Port 3/Port 4
Using the 2300-580-R Software with VectorStar ME7838D4
3-42
PN: 10410-00327 Rev. D
3659 0.8 mm Cal/Ver Kit & 2300-580-R PVS UG
3-17 High-Band Calibration – Port 3/Port 4
During this procedure, the high-band portion of the VNA is calibrated with a full 12-term SSST
(Short-Short-Short-Thru) calibration. In
Section 3-16 “Low-Band Calibration – Port 3/Port 4” on page 3-39
low-band portion of the VNA is calibrated with a full 12-term SOLT (Short-Open-Load-Thru) calibration.
Procedure
1.
The
0.8 mm Verification Install Adapter
dialog box appears.
a.
Install a F-F adapter on VNA Port 3 so that a Female connector is available to the operator.
b.
Install a F-M adapter on VNA Port 4 so that a Male connector is available to the operator.
c.
When ready to proceed, click
OK
.
2.
When ready to proceed, click
OK
. The
VNA Calibration: Step 1 of 5
dialog box appears.
a.
Connect the Male Cal Kit Isolation Device to VNA Port 3.
b.
Connect the Female Cal Kit Isolation Device to VNA Port 4.
c.
When ready to proceed, click
OK
.
Figure 3-59.
Anritsu 0.8 mm Verification Dialog
Figure 3-60.
VNA Calibration: Step 1 of 5 Dialog