Alarm and Warning Messages
This section lists the Error (Alarm / Warning) messages that can be triggered by
the system, to alert you to potential or impending problems with the analyzer.
Descriptions of the Alarms and Warnings, along with possible trigger condi-
tions, and corrective action to take to correct the Error, are also included.
Alarms
Alarm Condition / Description / Corrective Action
Cell T/C Failure
The Thermocouple is not connected or is open.
Trigger Conditions:
• Faulty
Thermocouple connection.
• Faulty
Thermocouple.
Corrective Action:
Take appropriate safety precautions, open the Analyzer door, and:
• Check the Thermocouple wiring for proper connections and inspect it for damage (cuts, nicks, burn marks,
etc.).
• Check the Thermocouple for a short or open circuit. Replace the Thermocouple if necessary.
Contact AMETEK
to verify operation before removing the Thermocouple.
Cold Junction Compensator Failure
The PCB temperature sensor has failed.
Trigger Conditions:
• This can be caused by faulty
electronics.
Corrective Action:
Take appropriate safety precautions, open the Analyzer door, and:
• Check the associated cables for proper connections and inspect them for damage (cuts, nicks, burn marks,
etc.).
• Test the PCB temperature sensor for proper operation. Use a multimeter to measure the resistance of the Tem-
perature Sensor and to test for an open circuit.
After this check has been made, cycle the analyzer AC power Off-On; this will reset the analyzer and may clear
any problems with the PCB.