PN 9000-280-VE, Rev B
Final Daisy Chain Connection Configuration Setting
When connecting multiple analyzers in a daisy chain configuration, enable the
communication termination resistor (Dip Switch
SW-5
) on the last analyzer in
the chain (farthest from the user interface).
SW-5
is located on the front of the
User Interface
board.
• SW-5 ON: Termination Enabled
• SW-5 OFF: Termination Disabled
The maximum distance between the AMEVision Display User Interface
and the analyzer is 1219.2 meters (4000 feet).
For instructions on how to work from the AMEVision Display User
Interface to configure and calibrate the analyzer, and to view analyzer
status readings, refer to the “AMEVision Display User Interface Installa-
tion and User Manual” (AMETEK PN 9000-165-VE).
2-32 | Thermox
®
WDG-V UOP ATEX-IECEx Analyzer