PN 9000-280-VE, Rev B
Remote Calibration Unit (RCU) Connections
This section discusses the electrical connections for the optional Remote Cali-
bration Unit (RCU):
Oxygen-Only Remote Calibration Unit Connections
Oxygen-Only Remote Calibration Unit (RCU) connections on the WDG-V UOP
Wiring
board, and their RCU connections, include:
ZERO GAS
WDG-V UOP Terminal
Z
to Pin
13
on RCU
ASPIRATOR
WDG-V UOP Terminal
A
to Pin
14
on RCU
O
2
SPAN
WDG-V UOP Terminal
S1
to Pin
15
on RCU
VALVE COMMON WDG-V UOP Terminal
V
to Pin
16
on RCU
Digital Input to Initiate Remote Calibration Unit
The digital input connections on the
Wiring
board allow you to initiate a
calibration from a remote location. For this option to work, you must have a
remote calibration unit. The system monitors the digital input, and when the
switch closes, it sends the system into an automatic calibration. The switch you
connect to this digital input must be a normally open switch.
Figure 2-10.
RCU Connections.
Installation and Start-Up
|
2-25