PN 9000-280-VE, Rev B
Alarm Condition / Description / Corrective Action
Last Flow Calibration Failed
The last Flow Sensor calibration failed.
Trigger Conditions:
• Plugged Sample System.
• Faulty O
2
Cell.
Corrective Action:
• Check the Sample System (and Inlet Probe) for plugging problems. Refer to the section “Plugged Plumbing
Check” for information on how to check for plugging problems.
• Check the O
2
Cell for proper operation.
Oxygen Calibration Required
O
2
Calibration is required.
Trigger Conditions:
• Last Calibration failed.
• Unit has never been calibrated.
Corrective Action:
• Restart the Calibration and allow sufficient time for it to complete its cycle.
3-32 | Thermox
®
WDG-V UOP ATEX-IECEx Analyzer