Alarm Condition / Description / Corrective Action
Analog Output 3 Out Of Range
The analog output value is saturated (high or low).
Trigger Conditions:
• Incorrect range setting.
Corrective Action:
From the AMEVision User Interface or PC Configurator Software:
• Check the range configured for Analog Output 3. Change it to the required range if necessary.
Last O
2
Span Calibration Failed
The last oxygen Span calibration failed.
Trigger Conditions:
• Span Calibration gas tank is empty.
• Elapsed Span calibration time too short.
• Faulty Calibration gas solenoid.
• Faulty O
2
Cell.
Corrective Action:
• Check the Span Calibration gas tank. Replace if necessary.
• Allow sufficient time to allow the Span Calibration gas to flow through the O
2
Sensor and to stabilize. Refer to
“Calibration/Aspirator Air Setup Checks” in this chapter for additional information.
• Check the Calibration gas solenoid for proper operation.
• Check the O
2
Cell for proper operation.
Last O
2
Zero Calibration Failed
The last oxygen Zero calibration failed.
Trigger Conditions:
• Zero Calibration gas tank is empty.
• Elapsed Zero calibration time too short.
• Faulty Calibration gas solenoid.
• Faulty O
2
Cell.
Corrective Action:
• Check the Zero Calibration gas tank. Replace if necessary.
• Allow sufficient time to allow the Zero Calibration gas to flow through the O
2
Sensor and to stabilize. Refer to
“Calibration/Aspirator Air Setup Checks” in this chapter for additional information.
• Check the Calibration gas solenoid for proper operation.
• Check the O
2
Cell for proper operation.
Maintenance and Troubleshooting
|