PN 9000-280-VE, Rev B
Alarm Condition / Description / Corrective Action
Box Temp Rise Failure
Box temperature failed to heat.
Trigger Conditions:
• This condition may be triggered upon analyzer power-up while the Box temperature is warming up to its
normal operating temperature, but will clear after the Cell has properly warmed up to its normal operating
temperature.
• Faulty Box Heater connection.
• Faulty RTD connection.
• Faulty Box Heater.
• Faulty RTD.
Corrective Action:
Take appropriate safety precautions, open the Analyzer door, and:
• Check the Box Heater and Box Temperature RTD wiring for proper connections and inspect them for damage
(cuts, nicks, burn marks, etc.).
• Check the Box Temperature RTD for a faulty circuit. Replace the RTD if necessary.
Contact AMETEK to verify
operation before removing the RTD.
• Check the Box Heater for proper operation. Replace the Heater if necessary.
Contact AMETEK to verify opera-
tion before removing the RTD.
Low Sample Flow
Low sample flow detected.
Trigger Conditions:
• This condition may be triggered upon analyzer power-up while the analyzer is warming up, but will clear after
the analyzer has properly warmed up to its normal operating temperature.
• Loss of Aspirator Air.
• Plugged Inlet Probe.
• Plugged Sample System.
Corrective Action:
• Check the Aspirator Air line for damage.
• Check the Aspirator Air supply to ensure it is adequate.
Take appropriate safety precautions, power down the Analyzer, and:
• Check the Sample System and Inlet Probe for plugging problems. Refer to the section “Plugged Plumbing
Check” for information on how to check for plugging problems.
Maintenance and Troubleshooting
|
3-27