PN 9000-280-VE, Rev B
Setting the Analyzer Address Using the Dip Switch (to
Enable Communications Between the Analyzer and the
AMEVision Display User Interface Unit)
The connection between the AMEVision Display User Interface and the
WDG-V UOP analyzer is a two-wire MODBUS RTU. A maximum of eight (8) ana-
lyzers can be connected to a single-user Display User Interface unit. Each con-
nected analyzer must have a unique address. The address on the WDG-V UOP
Analyzer is set using the dip switch on the front of the
Sensor
board. See
Figure 2-11 (and 2-14 for a close-up view of the dip switch).
Figure 2-14.
Dip switch settings on User
Interface PCB.
321 Switch
Address
001
1
010
2
011
3
100
4
1 = ON (UP)
0 = OFF (DOWN)
2-30 | Thermox
®
WDG-V UOP ATEX-IECEx Analyzer