Troubleshooting
SPECTRO ARCOS FHX22 — 18.07.2019
— 53
Malfunction
Cause
Remedy
Plasma flickers or
pulsates.
Moisture or deposits in the
aerosol tube of the plasma
torch.
Clean and dry the plasma torch
or replace it.
Aerosol pulsates (deposits
in nebulizer).
Clean the nebulizer.
Replace the pump tube.
Plasma too bright.
Power set too high.
Decrease power.
Oxygen in sample
introduction system.
Check for leaks and eliminate
these.
Plasma extinguishes
when nebulizer gas flow
is switched on.
Sample introduction tube
not connected to the
nebulizer.
Connect the tube and check
fitting
Oxygen in sample
introduction system.
Flush the system with argon
before restarting system.
Leak in the sample
introduction system.
Check for leaks and eliminate
these.
No channel visible in the
plasma.
Nebulizer gas flow setting
too low.
Optimize the nebulizer gas flow.
Nebulizer clogged.
Clean the nebulizer.
Leak in spray chamber
outlet.
Check for leaks and eliminate
these.
Error message:
Waste container is full.
Waste container full.
Remove the cover of the
container and properly dispose of
the content.
Sparkover at load coil.
Humidity.
Switch off the generator and use
a paper towel to dry the load coil
(if moisture is visible).
Dirt on the load coil.
Switch off the generator and
clean the load coil.
System information:
Wavelength calibration
failed.
Wrong sample or wrong
preflush time.
Use the correct sample. Check
the measurement parameters.
Repeat the wavelength
calibration.
Error message:
Network connection
No communication
between the PC and the
instrument.
Spectrometer not switched
on.
Address incorrect or used
for a different purpose.
Data cable not plugged in
or defective.
Check main switch.
Enter the correct IP address
(online help). Only qualified IT
staff are permitted to enter the
address.
Check cable.