4.2
Set-up
Caution! Measurement errors!
If the instrument is subject to vibrations and shocks, measurement errors
may occur during the measuring process.
The instrument must be installed at a vibration- and shock-free location.
Measurement errors result from temperature differences inside the
instrument.
The instrument must be installed at a location, where it cannot heat up
or cool down on one side only.
Observe the following information regarding the instrument's installation:
All connections are located on the right-hand side.
The instrument must be accessible from the front and from both sides for
maintenance purposes. Set up the instrument accordingly.
A table of corresponding dimensions and load-carrying capacity is required,
see page 15.
In addition to the space required by the SPECTRO ARCOS FHX22, some
space is needed for the gas supply, unless fixed gas connections are
available.
The SPECTRO ARCOS FHX22 is operated exclusively from the front and
from the right-hand side of the housing. Sufficient space should be provided
for the computer workstation (keyboard, monitor, printer, and sample changer,
if necessary) and operating personnel.
4.3
Ambient temperature
Air conditioning of the room is only required under extraordinary conditions.
Smaller room temperature fluctuations up to 3°C/h (5.4 F/h) are not accounted due
to the construction of the instrument. However, one-sided heating or cooling must
be avoided at all costs.
The specified values only apply to the recommended ambient temperature range.
Using the instrument beyond the recommended ambient conditions may lead to
poorer analytical results.