Maintenance
SPECTRO ARCOS FHX22 — 18.07.2019
— 45
2.
Insert a clean 'peek' sample
needle or similar type of object
(1) through the opening in the
back panel of the SPI (2).
3.
Carefully push the lens (3) out
on lens cleaning paper.
Fig. 23
4.
Clean the contaminated lens using lens cleaning paper or a lint-free soft cloth.
5.
Clean all the other parts with a lint-free cloth.
The lens and SPI are now clean.
6.9.3
Installing the lens and the SPI
In order to fit it, proceed as follows:
1.
Lay the PTFE ring (2) flat into
the SPI (1).
2.
Insert the lens (3) with the flat
side into the SPI.
The lens for the SPI (SOP
application) can be
distinguished by the straight
edge on two sides.
3.
Lay the second PTFE ring (4) on
top of it.
4.
Moderately tighten the light
channel sleeve (5) including the
light channel (6).
Fig. 24
5.
Insert the SPI into the instrument following the reverse order used for the
disassembly procedure.
6.
Install the plasma torch and adapter, see page 32.
The lens and SPI are now installed.
2
1
3
1
2
3
4
5
6